Impact of surface porosity and topography on the mechanical behavior of high strength biomedical polymers.

Nathan T Evans1, Cameron W Irvin2, David L Safranski3, Ken Gall4.   

Abstract

The ability to control the surface topography of orthopedic implant materials is desired to improve osseointegration but is often at the expense of mechanical performance in load bearing environments. Here we investigate the effects of surface modifications, roughness and porosity, on the mechanical properties of a set of polymers with diverse chemistry and structure. Both roughness and surface porosity resulted in samples with lower strength, failure strain and fatigue life due to stress concentrations at the surface; however, the decrease in ductility and fatigue strength were greater than the decrease in monotonic strength. The fatigue properties of the injection molded polymers did not correlate with yield strength as would be traditionally observed in metals. Rather, the fatigue properties and the capacity to maintain properties with the introduction of surface porosity correlated with the fracture toughness of the polymers. Polymer structure impacted the materials relative capacity to maintain monotonic and cyclic properties in the face of surface texture and porosity. Generally, amorphous polymers with large ratios of upper to lower yield points demonstrated a more significant drop in ductility and fatigue strength with the introduction of porosity compared to crystalline polymers with smaller ratios in their upper to lower yield strength. The latter materials have more effective dissipation mechanisms to minimize the impact of surface porosity on both monotonic and cyclic damage.
Copyright © 2016 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Notch; Osseointegration; Polyetheretherketone; Surface porous; Topography

Mesh:

Substances:

Year:  2016        PMID: 26986085     DOI: 10.1016/j.jmbbm.2016.02.033

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  2 in total

1.  Getting PEEK to Stick to Bone: The Development of Porous PEEK for Interbody Fusion Devices.

Authors:  F Brennan Torstrick; David L Safranski; J Kenneth Burkus; James L Chappuis; Christopher S D Lee; Robert E Guldberg; Ken Gall; Kathryn E Smith
Journal:  Tech Orthop       Date:  2017-09-01

2.  Attachment, Proliferation, and Morphological Properties of Human Dermal Fibroblasts on Ovine Tendon Collagen Scaffolds: A Comparative Study.

Authors:  Fauzi Mh Busra; Yogeswaran Lokanathan; Masrina Mohd Nadzir; Aminuddin Saim; Ruszymah Bt Hj Idrus; Shiplu Roy Chowdhury
Journal:  Malays J Med Sci       Date:  2017-04-14
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.