Detecting MLC errors in stereotactic radiotherapy plans with a liquid filled ionization chamber array.

Patrick O'Connor1, Venkatakrisnan Seshadri2, Paul Charles2,3.   

Abstract

Quality assurance of stereotactic radiotherapy demands the use of equipment with the highest resolution and sensitivity available. This study examines the sensitivity of a commercially available liquid-filled ionization chamber array--the Octavius 1000 SRS (PTW, Frieburg, Germany) for detecting small (sub-millimetre) multi-leaf collimator (MLC) alignment errors in static square fields (side length 16-40 mm). Furthermore, the effectiveness of detecting small MLC errors in clinical stereotactic radiotherapy patient plans using the device was also evaluated. The commonly used gamma pass rate metric (of the measurements compared with treatment planning system generated results) was used. The gamma pass rates were then evaluated as a function of MLC position error (MLC error size 0.1-2.5 mm). The detector array exhibited a drop in pass rate between plans without error and those which had MLC errors induced. For example a drop in pass rate of 4.5% (gamma criteria 3%, 1 mm) was observed when a 0.8 mm error was introduced into a 16 mm square field. Furthermore the drop in pass rate increased as the MLC position error increased. This study showed that the Octavius 1000 SRS array could be a useful tool for applications requiring the detection of small geometric delivery uncertainties.

Entities:  

Keywords:  Dosimetry; Octavius 1000 SRS; Patient specific QA; Stereotactic QA

Mesh:

Year:  2016        PMID: 26979835     DOI: 10.1007/s13246-016-0421-6

Source DB:  PubMed          Journal:  Australas Phys Eng Sci Med        ISSN: 0158-9938            Impact factor:   1.430


  2 in total

1.  Dosimetric comparison of different treatment modalities for stereotactic radiotherapy.

Authors:  Shih-Ming Hsu; Yuan-Chun Lai; Chien-Chung Jeng; Chia-Ying Tseng
Journal:  Radiat Oncol       Date:  2017-09-16       Impact factor: 3.481

2.  The impact of isocentric shifts on delivery accuracy during the irradiation of small cerebral targets-Quantification and possible corrections.

Authors:  Linda J Wack; Florian Exner; Sonja Wegener; Otto A Sauer
Journal:  J Appl Clin Med Phys       Date:  2020-03-20       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.