| Literature DB >> 2697515 |
Abstract
A new CT stereotactic device named version WL-86 was designed and manufactured by the authors. Phantom puncture and cadaver puncture tests showed that its accuracy is less than 1 mm. We have also made a new type of biopsy needle--the rotary and translational cutting biopsy needle. A definite pathological diagnosis was made in 93.7% (30/32) of patients by this technique: The practicability of the device, the indications and clinical value of CT-guided stereotactic biopsy were discussed.Entities:
Mesh:
Year: 1989 PMID: 2697515
Source DB: PubMed Journal: Zhonghua Fang She Xue Za Zhi ISSN: 1005-1201