Aerosol Delivery Using Jet Nebulizer and Vibrating Mesh Nebulizer During High Frequency Oscillatory Ventilation: An In Vitro Comparison.

Tien-Pei Fang1,2, Hui-Ling Lin3, Shu-Hua Chiu1, Szu-Hui Wang1, Robert M DiBlasi4, Ying-Huang Tsai5, James B Fink6.   

Abstract

BACKGROUND: High frequency oscillatory ventilation (HFOV) is used in critically ill patients with severe hypoxemic respiratory failure. The purpose of this in vitro study was to determine the efficiency of aerosol delivery with different lung parameters during simulated neonatal, pediatric, and adult HFOV.
METHODS: Sensormedics 3100A/B ventilators were used to deliver infant, pediatric, and adult HFOV. Two types of aerosol generators were chosen for testing: 1) a continuous jet nebulizer (JN) with a unit-dose of 5.0 mg/2.5 mL salbutamol sulfate diluted into 4 mL, and 2) a vibrating mesh nebulizer (VMN) with salbutamol sulfate were run to completion of aerosol generation. Both aerosol devices were placed 1) between the ventilator circuit and the endotracheal tube (ETT) (proximal position); and 2) at the inlet of the heated humidifier (distal position) (n = 5). Drug was collected on a bacterial filter placed distal to the ETT, and the drug eluted and analyzed with a UV Spectrophotometer at 276 nm. T- test and ANOVA tests were used for comparison (p < 0.05).
RESULTS: The inhaled drug delivered by JN was 0%-0.6% of the nominal dose when placed at distal position, and 0%-3% at proximal position (p < 0.01), while the VMN was 0%-0.5% at distal and 8.6%-22.7% at proximal position (p < 0.01). Aerosol delivery during HFOV was greater with adult settings than pediatric and infant settings with VMN and JN (22.7%, 8.6%, and 17.4% respectively, p < 0.01). When the aerosol delivery device was placed at the distal position, negligible drug mass was observed (<0.5%), regardless of the nebulizer device used.
CONCLUSIONS: During HFOV, aerosol delivery with the nebulizer placed at proximal was greater than placement distal from the ETT, with VMN delivering more drug than JN. The inhaled drug was delivery correlated positively with ETT size, MAP, and bias flow, and inversely proportional to power settings.

Entities:  

Keywords:  aerosol delivery; breathing patterns; high frequency oscillatory ventilation; jet nebulizer; vibrating mesh nebulizer

Year:  2016        PMID: 26974175     DOI: 10.1089/jamp.2015.1265

Source DB:  PubMed          Journal:  J Aerosol Med Pulm Drug Deliv        ISSN: 1941-2711            Impact factor:   2.849


  4 in total

Review 1.  Aerosol delivery via invasive ventilation: a narrative review.

Authors:  Hui-Ling Lin; James B Fink; Huiqing Ge
Journal:  Ann Transl Med       Date:  2021-04

2.  Inhaled Treprostinil Drug Delivery During Mechanical Ventilation and Spontaneous Breathing Using Two Different Nebulizers.

Authors:  Donna K Parker; Shuijie Shen; Jiang Zheng; D Dunbar Ivy; Dave N Crotwell; Justin C Hotz; Robert M DiBlasi
Journal:  Pediatr Crit Care Med       Date:  2017-06       Impact factor: 3.624

3.  In vitro evaluation of disposable transport ventilators with combination aerosol therapy.

Authors:  Marc Mac Giolla Eain; Andrew O'Sullivan; Mary Joyce; Ronan MacLoughlin
Journal:  BMJ Open Respir Res       Date:  2021-03

4.  Quantitative shadowgraphy of aerosol and droplet creation during oscillatory motion of the microkeratome amid COVID-19 and other infectious diseases.

Authors:  Pooja Khamar; Rohit Shetty; Nikhil Balakrishnan; Prasenjit Kabi; Durbar Roy; Saptarshi Basu; Abhijit Sinha Roy
Journal:  J Cataract Refract Surg       Date:  2020-10       Impact factor: 3.528

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.