Abstract
Quantitative fluorescence turn-on and colorimetric detection of glutathione (GSH) with rapid speed, low cost have attained much attention. Herein, we developed a sensitive fluorescence turn-on and colorimetric sensor for GSH based on the inner-filter effect (IFE), which is the first time to select oxTMB and TMPyP as the IFE absorber and fluorophore pair, respectively. The absorption band of oxTMB matches well with the emission band of TMPyP in the IFE-based fluorescent assay. In the absence of GSH, the absorption peak of oxTMB at 652nm significantly overlaps with the emission of TMPyP, resulting in the efficient IFE and inhibition of the fluorescence of TMPyP. In the presence of GSH, the absorption intensity at 652nm decreases, generating the recovery of the fluorescence of TMPyP. Therefore, this approach is demonstrated to be a novel candidate for detection of GSH, with high sensitivity and selectivity. The linear dynamic range for the concentrations of GSH is between 0.1μM to 20μM along with a limit of detection (LOD) of about 30nM (calculated LOD as 3σ/slope). Finally, this novel sensor was successfully applied for GSH detection in fetal calf serum, and satisfactory recovery was achieved.Entities:
Keywords: Fluorescence turn-on and colorimetric assay; Glutathione; Inner-filter effect; TMB; TMPyP
Mesh:
Substances:
Year: 2016 PMID: 26971272 DOI: 10.1016/j.bios.2016.02.068
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618