Hemoglobin assessment: precision and practicability evaluated in the Netherlands-the HAPPEN study.

A Mireille Baart1, Wim L A M de Kort1, Katja van den Hurk1, Pieternel C M Pasker-de Jong1.   

Abstract

BACKGROUND: Most donors do not like the invasive hemoglobin (Hb) measurement in fingerstick capillary samples. Alternative noninvasive devices have recently become available. In this study we assessed the precision and practicability of noninvasive devices and compared them with measurements in capillary and venous samples. STUDY DESIGN AND METHODS: A total of 2589 blood donors were included in the study. Hb levels were measured with the noninvasive devices NBM 200 or Haemospect and compared with values obtained with the invasive devices HemoCue 201 and HemoCue 301 and with venous values. Precision was evaluated in terms of sensitivity and specificity for detection of Hb levels below the cutoff level for donation, taking venous values as "reference values."
RESULTS: Hb levels measured with both invasive and noninvasive devices differed significantly from venous measurements. Measurements with invasive devices correlated stronger with venous values and were more precise. Sensitivity ranged from 30.1% (HemoCue 201) to 3.5% (Haemospect) in men and from 36.6% (HemoCue 201) to 10.8% (Haemospect) in women. Specificity ranged from 99.8% (HemoCue 201) to 95.5% (Haemospect) in men and from 98.9% (HemoCue 201) to 94.6% (NBM 200) in women. Most donors preferred the noninvasive measurement.
CONCLUSION: Measurements with invasive devices showed stronger correlation and better agreement with venous values than noninvasive devices. Sensitivity was low for all measurement devices, particularly the noninvasive ones. In terms of precision invasive measurement would be preferred for donor screening, although a majority of donors preferred the noninvasive measurement.
© 2016 AABB.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26968697     DOI: 10.1111/trf.13546

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  7 in total

1.  Validation of Point of Care Hemoglobin Estimation Among Pregnant Women Using Digital Hemoglobinometers (HemoCue 301 and HemoCue 201+) as Compared with Auto-Analyzer.

Authors:  Kapil Yadav; Shashi Kant; Gomathi Ramaswamy; Farhad Ahamed; Olivia Marie Jacob; Himani Vyas; Ravneet Kaur; Sumit Malhotra; Partha Haldar
Journal:  Indian J Hematol Blood Transfus       Date:  2019-10-01       Impact factor: 0.900

Review 2.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

Review 3.  Emerging point-of-care technologies for anemia detection.

Authors:  Ran An; Yuning Huang; Yuncheng Man; Russell W Valentine; Erdem Kucukal; Utku Goreke; Zoe Sekyonda; Connie Piccone; Amma Owusu-Ansah; Sanjay Ahuja; Jane A Little; Umut A Gurkan
Journal:  Lab Chip       Date:  2021-05-18       Impact factor: 6.799

Review 4.  Techniques used for the screening of hemoglobin levels in blood donors: current insights and future directions.

Authors:  Rajendra Chaudhary; Anju Dubey; Atul Sonker
Journal:  J Blood Med       Date:  2017-07-03

5.  Hemoglobin concentration, total hemoglobin mass and plasma volume in patients: implications for anemia.

Authors:  James M Otto; James O M Plumb; Eleri Clissold; Shriya B Kumar; Denis J Wakeham; Walter Schmidt; Michael P W Grocott; Toby Richards; Hugh E Montgomery
Journal:  Haematologica       Date:  2017-06-08       Impact factor: 9.941

6.  Comparison of four methods to measure haemoglobin concentrations in whole blood donors (COMPARE): A diagnostic accuracy study.

Authors:  Steven Bell; Michael Sweeting; Anna Ramond; Ryan Chung; Stephen Kaptoge; Matthew Walker; Thomas Bolton; Jennifer Sambrook; Carmel Moore; Amy McMahon; Sarah Fahle; Donna Cullen; Susan Mehenny; Angela M Wood; Jane Armitage; Willem H Ouwehand; Gail Miflin; David J Roberts; John Danesh; Emanuele Di Angelantonio
Journal:  Transfus Med       Date:  2020-12-20       Impact factor: 2.019

7.  Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India.

Authors:  Kapil Yadav; Shashi Kant; Gomathi Ramaswamy; Farhad Ahamed; Kashish Vohra
Journal:  Indian J Community Med       Date:  2020-10-28
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.