Abstract
This paper presents an ultra HPLC/electrospray ionization-tandem MS method to determine pesticides in wine. We adopted the quick, easy, cheap, effective, rugged, and safe (QuEChERs) method for extraction and used core-shell column to achieve ultra-HPLC to develop and validate a simple and fast method to analyze 187 pesticide residues in red and white wine samples. Pesticide residues were extracted from wine samples using QuEChERS. Ultra HPLC/electrospray ionization-tandem MS quantification was achieved using matrix-matched standard calibration curves with isotopically labeled standards or a chemical analogue as internal standards with an analytical range from 5.0 to 500.0 μg/L. The method performance characteristics that included overall recovery, intermediate precision, and measurement uncertainty were evaluated according to a nested experimental design. Generally, 98.4% (in red wine) and 96.8% (in white wine) of the pesticides had recoveries between 71 and 120%; 98.9% (in red wine) and 99.5% (in white wine) of the pesticides had the intermediate precision ≤20%; and 99.5% (in red wine) and 98.4% (in white wine) of the pesticides had measurement uncertainty ≤50%.Mesh:
Substances:
Year: 2016 PMID: 26965037 DOI: 10.5740/jaoacint.15-0265