| Literature DB >> 26959840 |
Dananjaya K Appulage1, Evelyn H Wang1, Frances Carroll2, Kevin A Schug1.
Abstract
There are various reversed-phase stationary phases that offer significant differences in selectivity and retention. To investigate different reversed-phase stationary phases (aqueous stable C18 , biphenyl, pentafluorophenyl propyl, and polar-embedded alkyl) in an automated fashion, commercial software and associated hardware for mobile phase and column selection were used in conjunction with liquid chromatography and a triple quadrupole mass spectrometer detector. A model analyte mixture was prepared using a combination of standards from varying classes of analytes (including drugs, drugs of abuse, amino acids, nicotine, and nicotine-like compounds). Chromatographic results revealed diverse variations in selectivity and peak shape. Differences in the elution order of analytes on the polar-embedded alkyl phase for several analytes showed distinct selectivity differences compared to the aqueous C18 phase. The electron-rich pentafluorophenyl propyl phase showed unique selectivity toward protonated amines. The biphenyl phase provided further changes in selectivity relative to C18 with a methanolic phase, but it behaved very similarly to a C18 when an acetonitrile-based mobile phase was evaluated. This study shows the value of rapid column screening as an alternative to excessive mobile phase variation to obtain suitable chromatographic settings for analyte separation.Entities:
Keywords: Automated method scouting; Column switching; Drugs of abuse; Selectivity
Mesh:
Substances:
Year: 2016 PMID: 26959840 DOI: 10.1002/jssc.201600131
Source DB: PubMed Journal: J Sep Sci ISSN: 1615-9306 Impact factor: 3.645