| Literature DB >> 26959610 |
Michael J Parkes1,2, Stuart Green3, Andrea M Stevens4, Sophia Parveen3, Rebecca Stephens3, Thomas H Clutton-Brock1,5.
Abstract
OBJECTIVE: Variability in the breathing pattern of patients with cancer during radiotherapy requires mitigation, including enlargement of the planned treatment field, treatment gating and breathing guidance interventions. Here, we provide the first demonstration of how easy it is to mechanically ventilate patients with breast cancer while fully conscious and without sedation, and we quantify the resulting reduction in the variability of breathing.Entities:
Mesh:
Year: 2016 PMID: 26959610 PMCID: PMC5258146 DOI: 10.1259/bjr.20150741
Source DB: PubMed Journal: Br J Radiol ISSN: 0007-1285 Impact factor: 3.039
Figure 1.Equipment on the patient during mechanical ventilation in the simulator room. The figure shows a patient lying supine on a breast board, with blood pressure (BP) and oxygen saturation (SpO2) measured with non-invasive monitors on the fingers, a three-lead electrocardiogram (ECG) to measure heart rate and airway pressure and exhaled partial pressure of carbon dioxide (PCO2) measured in the face mask. The mechanical ventilator drives breathing via the face mask. The Osiris measures the movement of the chest with markers positioned at the centre of the chest, on the left breast and with right and left lateral markers oriented at 90° to the horizontal. The hypothetical X-ray target is indicated by the dotted circle. The Osiris surface-image mapping system was obtained from Qados Ltd, Sandhurst, UK.
Figure 2.Mechanical ventilation regularizes breathing and breast movement in Patient 1. (a) Airway pressure during 2 min of spontaneous (irregular) breathing. Pressure measurements are uncalibrated [in arbitrary (arb.) units]. Spontaneous breaths in the mask cause negative pressure waves (downwards in the figure); but, for comfort, the ventilator then added approximately 10 cm water (approximately 10 mbar) of inspiratory assist, so each spontaneous breath continues as a positive pressure wave (upwards in the figure). For spontaneous breathing, the mean breathing frequency with its variability [±standard deviation (SD)%] within this patient in breaths per minute, inflation volume with its variability (±SD%) in arb. units seconds, the recording period duration (minutes) and number (#) of breaths during this period are indicated. (b) Airway pressure during 2 min of mechanical ventilation. The regular frequency and pressure amplitude indicate that the patient is passive (i.e. has allowed the ventilator to take over their breathing). For mechanical ventilation, the mean breathing frequency (f) with its variability (±SD%) within this patient in breaths per minute, inflation volume with its variability (±SD%) in arb. units seconds, the recording period duration (minutes) and number (#) of breaths during this period are indicated. (c) Left breast anteroposterior (ant.-post.) movement during 50 s of mechanical ventilation. The regular frequency and movement amplitude of the left breast confirms that the patient is passive and demonstrates how predictable is the breast movement during mechanical ventilation. The mean displacement (mm), variability (var) in peak and trough position (mm) and peak and trough drift (mm min−1) of the left breast marker are indicated. (d) Airway pressure during the same 50 s of mechanical ventilation as in (c). This demonstrates the correspondence between airway pressure and breast movement during mechanical ventilation and therefore that inflation volume is the same for each mechanically induced breath. For mechanical ventilation, the mean breathing frequency with its variability (±SD%) within this patient in breaths per minute, inflation volume with its variability (±SD%) in arb. units seconds, the recording period duration (minutes) and number (#) of breaths during this period are indicated.
Figure 5.Mechanical ventilation regularizes breathing and breast movement in Patient 4. For descriptions on (a)–(d), refer to the legend of Figure 2.