The efficacy and safety of a novel posterior scleral reinforcement device in rabbits.

Yongguang Yuan1, Yao Zong1, Qishan Zheng1, Garrett Qian2, Xiaobin Qian1, Yujie Li1, Wanwen Shao1, Qianying Gao3.   

Abstract

PURPOSE: To evaluate the efficacy and safety of posterior scleral reinforcement (PSR) device for myopia suppression in rabbits' eyes.
METHODS: PSR surgery was performed on the normal 12 8-week-old New Zealand white rabbits' right eyes. To determine efficacy of the device, ophthalmic examination would be taken at pre-operation and post-operation (1 week, 1 month, 3 months, 6 months, and 1 year), such as A-ultrasound, diopter and B-ultrasound. Evaluation of safety were based on the following indicators: intraocular pressure (IOP), slit lamp, fundus photography, fundus fluorescein angiography and pathological examination after surgery. The efficacy and safety of PSR device were evaluated by comparison (treated eyes and contralateral eyes) of pre and post-operation.
RESULTS: The novel PSR device could significantly shorten axial length (preoperative axial length: 16.36 ± 0.14 mm, postoperative 1 week, 1 month, 3 months, 6 months and 1 year axial lengths: 15.03 ± 0.28 mm, 15.23 ± 0.32 mm, 15.39 ± 0.31 mm, 15.45 ± 0.22 mm and 15.45 ± 0.22 mm; P=0.00037<0.001) in the treated eyes (right eyes) after surgery. At different postoperative time points, the B-ultrasound images showed that the PSR located in appropriate position and supported the posterior sclera very well. At the same time, IOP of treated eyes kept a relatively stable level (preoperative IOP: 12.56 ± 2.01 mmHg, postoperative IOP: ranging from 11.33 ± 1.23 mmHg to 13.44 ± 2.19 mmHg, P>0.05) post-operation 1 year. During observation period, there was no significant inflammatory reaction and complications such as anterior chamber flare, empyema, endophthalmitis, vitreous hemorrhage, retina detachment and retinal choroid neovascularization by slit lamp, fundus photography and fundus fluorescein angiography. In addition, there were no pathologic changes be found by comparison treated eyes group and contralateral group eyes based on pathological examinations.
CONCLUSIONS: In vivo study, effectively and safely, the novel PSR device can inhibit rabbits' axial length elongation during postoperative 1 year. This study demonstrates that this novel PSR could be a potential treatment approach for myopia.
Copyright © 2016 The Authors. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Axial length; Myopia; PSR; Retina

Mesh:

Year:  2016        PMID: 26952419     DOI: 10.1016/j.msec.2015.12.046

Source DB:  PubMed          Journal:  Mater Sci Eng C Mater Biol Appl        ISSN: 0928-4931            Impact factor:   7.328


  1 in total

1.  Study of retina and choroid biological parameters of rhesus monkeys eyes on scleral collagen cross-linking by riboflavin and ultraviolet A.

Authors:  Mingshen Sun; Fengju Zhang; Bowen Ouyang; Mengmeng Wang; Yu Li; Xuan Jiao; Li Zhang; Ningli Wang
Journal:  PLoS One       Date:  2018-02-08       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.