Effectiveness of the SYSTEM 1E Liquid Chemical Sterilant Processing System for reprocessing duodenoscopes.

Gerald McDonnell1, Michele Ehrman2, Sara Kiess1.   

Abstract

BACKGROUND: A troubling number of health care-acquired infection outbreaks and transmission events, some involving highly resistant microbial pathogens and resulting in serious patient outcomes, have been traced to reusable, high-level disinfected duodenoscopes in the United States. The Food and Drug Administration (FDA) requested a study be conducted to verify liquid chemical sterilization efficacy of SYSTEM 1E(®) Liquid Chemical Sterilant Processing System (STERIS Corporation, Mentor, OH) with varied duodenoscope designs under especially arduous conditions. Here, we describe the system's performance under worst case SYSTEM 1E(®) processing conditions.
METHODS: The test protocol challenged the system's performance by running a fractional cycle to evaluate reduction of recoverable test spores from heavily contaminated endoscopes, including all channels and each distal tip, under worst case SYSTEM 1E(®) processing conditions.
RESULTS: All devices were successfully liquid chemically sterilized, showing greater than a 6 log10 reduction of Geobacillus stearothermophilus spores at every inoculation site of each duodenoscope tested, in less than half the exposure time of the standard cycle.
CONCLUSIONS: The successful outcome of the additional efficacy testing reported here indicates that the SYSTEM 1E(®) is an effective low-temperature liquid chemical sterilization method for duodenoscopes and other critical and semicritical devices. It offers a fast, safe, convenient processing alternative while providing the assurance of a system expressly tested and cleared to achieve liquid chemical sterilization of specific validated duodenoscope models.
Copyright © 2016 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  GI endoscope reprocessing; Liquid chemical sterilization; Low-temperature sterilization; Semicritical device; Sterilization

Mesh:

Substances:

Year:  2016        PMID: 26948402     DOI: 10.1016/j.ajic.2016.01.008

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  3 in total

Review 1.  Duodenoscope-Associated Infections: Update on an Emerging Problem.

Authors:  M Rubayat Rahman; Abhilash Perisetti; Roxana Coman; Pardeep Bansal; Rajiv Chhabra; Hemant Goyal
Journal:  Dig Dis Sci       Date:  2018-12-19       Impact factor: 3.199

Review 2.  Recent Advances in Endoscope Disinfection: Where Do We Stand in the COVID era?

Authors:  Tiffany Chua; Nasir Halim; Sofiya Reicher
Journal:  Tech Innov Gastrointest Endosc       Date:  2020-10-19

Review 3.  A narrative review on current duodenoscope reprocessing techniques and novel developments.

Authors:  Maarten Heuvelmans; Herman F Wunderink; Henny C van der Mei; Jan F Monkelbaan
Journal:  Antimicrob Resist Infect Control       Date:  2021-12-23       Impact factor: 4.887

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.