Literature DB >> 26930674

Real-Time Model-Based Fault Detection of Continuous Glucose Sensor Measurements.

Kamuran Turksoy, Anirban Roy, Ali Cinar.   

Abstract

OBJECTIVE: Faults in subcutaneous glucose concentration readings with a continuous glucose monitoring (CGM) may affect the computation of insulin infusion rates that can lead to hypoglycemia or hyperglycemia in artificial pancreas control systems for patients with type 1 diabetes (T1D).
METHODS: Multivariable statistical monitoring methods are proposed for detection of faults in glucose concentration values reported by a subcutaneous glucose sensor. A nonlinear first principle glucose/insulin/meal dynamic model is developed. An unscented Kalman filter is used for state and parameter estimation of the nonlinear model. Principal component analysis models are developed and used for detection of dynamic changes. K-nearest neighbor classification algorithm is used for diagnosis of faults. Data from 51 subjects are used to assess the performance of the algorithm.
RESULTS: The results indicate that the proposed algorithm works successfully with 84.2% sensitivity. Overall, 155 (out of 184) of the CGM failures are detected with a 2.8-min average detection time.
CONCLUSION: A novel algorithm that integrates data-driven and model-based methods is developed. The proposed method is able to detect CGM failures with a high rate of success. SIGNIFICANCE: The proposed fault detection algorithm can decrease the effects of faults on insulin infusion rates and reduce the potential for hypo- or hyperglycemia for patients with T1D.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26930674     DOI: 10.1109/TBME.2016.2535412

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  3 in total

1.  Hypoglycemia Detection and Carbohydrate Suggestion in an Artificial Pancreas.

Authors:  Kamuran Turksoy; Jennifer Kilkus; Iman Hajizadeh; Sediqeh Samadi; Jianyuan Feng; Mert Sevil; Caterina Lazaro; Nicole Frantz; Elizabeth Littlejohn; Ali Cinar
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

Review 2.  Artificial Intelligence for Diabetes Management and Decision Support: Literature Review.

Authors:  Ivan Contreras; Josep Vehi
Journal:  J Med Internet Res       Date:  2018-05-30       Impact factor: 5.428

Review 3.  Continuous Glucose Monitoring Sensors: Past, Present and Future Algorithmic Challenges.

Authors:  Andrea Facchinetti
Journal:  Sensors (Basel)       Date:  2016-12-09       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.