Literature DB >> 26923497

Metal Ion Levels Are Not Correlated With Histopathology of Adverse Local Tissue Reactions in Taper Corrosion of Total Hip Arthroplasty.

Ming Han Lincoln Liow1, Kenneth L Urish1, Frederic I Preffer2, Gunnlaugur P Nielson2, Young-Min Kwon1.   

Abstract

BACKGROUND: The underlying biological mechanism in the formation of adverse local tissue reaction in taper corrosion of total hip arthroplasty (THA) remains unknown. This study evaluated whether there was a dose-dependent relationship between metal ion levels, intraoperative tissue damage and ALVAL (aseptic lymphocyte-dominated vasculitis-associated lesion) scores in dual taper THA patients who underwent revisions for taper corrosion.
METHODS: We performed a retrospective review of 31 dual taper THA patients who underwent revision surgery from May 2013 to October 2013. Preoperative serum metal ion levels, intraoperative tissue damage grading, and ALVAL scores were reviewed. Multivariate analysis was performed to determine if an association existed between metal ion levels, intraoperative tissue damage, and ALVAL scores.
RESULTS: Findings consistent with adverse local tissue reaction were found in all cases. We noted 10 patients with low, 8 with moderate, and 13 with high ALVAL scores, respectively. For intraoperative tissue damage, we recorded 2 (grade 1), 22 (grade 2) and 7 (grade 3) cases. Preoperatively, there was preferential elevation of serum cobalt (3.8 ng/mL, 2.3-17.0) compared to serum chromium (1.0 ng/mL, 0.2-5.8). There was no correlation between preoperative metal ion levels and intraoperative tissue damage (R = -0.06, P = .74) or ALVAL scores (R = -0.04, P = .481). There was also no correlation between intraoperative tissue damage and ALVAL score (R = -0.06, P = .73).
CONCLUSION: There was no significant correlation between ALVAL scores and prerevision surgery metal ion levels or intraoperative tissue damage, suggesting that the biological mechanism of histologic morphology cannot be solely attributed to elevated metal ion levels and is likely multifactorial, reflecting a complex interplay between implant and patient factors.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  ALVAL score; histopathology; intraoperative tissue damage; metal ion levels; taper corrosion; total hip arthroplasty

Mesh:

Substances:

Year:  2016        PMID: 26923497     DOI: 10.1016/j.arth.2016.01.041

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  13 in total

1.  Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.

Authors:  Kenneth L Urish; Brian R Hamlin; Anton Y Plakseychuk; Timothy J Levison; Genymphas B Higgs; Steven M Kurtz; Anthony M DiGioia
Journal:  J Arthroplasty       Date:  2017-04-08       Impact factor: 4.757

Review 2.  Metal-on-metal total hip arthroplasty: risk factors for pseudotumours and clinical systematic evaluation.

Authors:  Ming Han Lincoln Liow; Young-Min Kwon
Journal:  Int Orthop       Date:  2016-10-20       Impact factor: 3.075

3.  The influence of HLA genotype on the development of metal hypersensitivity following joint replacement.

Authors:  David J Langton; Rohan M Bhalekar; Thomas J Joyce; Stephen P Rushton; Benjamin J Wainwright; Matthew E Nargol; Nish Shyam; Benedicte A Lie; Moreica B Pabbruwe; Alan J Stewart; Susan Waller; Shonali Natu; Renne Ren; Rachelle Hornick; Rebecca Darlay; Edwin P Su; Antoni V F Nargol
Journal:  Commun Med (Lond)       Date:  2022-06-24

4.  Metal-on-Metal Hip Joint Prostheses: a Retrospective Case Series Investigating the Association of Systemic Toxicity with Serum Cobalt and Chromium Concentrations.

Authors:  James H Ho; Jerrold B Leikin; Paul I Dargan; John R H Archer; David M Wood; Jeffrey Brent
Journal:  J Med Toxicol       Date:  2017-09-12

5.  Outcome of revision surgery for adverse local tissue reactions in patients with recalled total hip arthroplasty.

Authors:  Christian Klemt; Sakkadech Limmahakhun; Georges Bounajem; Christopher M Melnic; Michael J Harvey; Young-Min Kwon
Journal:  Arch Orthop Trauma Surg       Date:  2021-04-09       Impact factor: 2.928

6.  Analysis of bearing wear, whole blood and synovial fluid metal ion concentrations and histopathological findings in patients with failed ASR hip resurfacings.

Authors:  Lari Lehtovirta; Aleksi Reito; Jyrki Parkkinen; Harry Hothi; Johann Henckel; Alister Hart; Antti Eskelinen
Journal:  BMC Musculoskelet Disord       Date:  2017-12-11       Impact factor: 2.362

7.  Activity-related Changes in Cobalt Levels in a Total Hip with a Modular Femoral Stem: A Case Report.

Authors:  Michael S Reich; Richard H Walker
Journal:  J Orthop Case Rep       Date:  2018 Jan-Feb

Review 8.  Adverse reaction to metal debris due to fretting corrosion between the acetabular components of modular dual-mobility constructs in total hip replacement: a systematic review and meta-analysis.

Authors:  Jonathan M R French; Paul Bramley; Sean Scattergood; Nemandra A Sandiford
Journal:  EFORT Open Rev       Date:  2021-05-04

9.  Association between periprosthetic tissue metal content, whole blood and synovial fluid metal ion levels and histopathological findings in patients with failed metal-on-metal hip replacement.

Authors:  Lari Lehtovirta; Aleksi Reito; Jyrki Parkkinen; Sirpa Peräniemi; Jouko Vepsäläinen; Antti Eskelinen
Journal:  PLoS One       Date:  2018-05-16       Impact factor: 3.240

10.  Retrieval Findings of Recalled Dual-Taper Hips.

Authors:  Anna Di Laura; Harry S Hothi; Johann Henckel; Young-Min Kwon; John A Skinner; Alister J Hart
Journal:  J Bone Joint Surg Am       Date:  2018-10-03       Impact factor: 5.284