Literature DB >> 26920190

Inappropriate Implantable Cardioverter-Defibrillator Shock from QRS Double Counting in the Setting of Hyperkalemia.

Akash Dadlani1, Jonathan W Dukes1, Nitish Badhwar2.   

Abstract

This case shows the complexity of arrhythmia management in patients with implantable cardioverter-defibrillators (ICDs) who present with hyperkalemia. In order to prevent inappropriate ICD shock, consideration should be given to the suspension of ICD therapies while intensive care treatment of extreme electrolyte derangements is being pursued. Patients in these setting should be closely monitored until their electrocardiograms have normalized, after which the device can safely be reactivated.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Hyperkalemia; ICD; Inappropriate shock

Mesh:

Year:  2016        PMID: 26920190     DOI: 10.1016/j.ccep.2015.10.024

Source DB:  PubMed          Journal:  Card Electrophysiol Clin        ISSN: 1877-9182


  1 in total

1.  A case of pseudo-appropriate shock.

Authors:  Neal J Ferrick; John D Fisher; Jay N Gross; Andrew K Krumerman
Journal:  HeartRhythm Case Rep       Date:  2020-03-13
  1 in total