BACKGROUND AND
OBJECTIVE: Body sensors are finding increasing applications in the self-monitoring for health-care and in the remote surveillance of sensitive people. The physiological data to be sampled can be non-stationary, with bursts of high amplitude and frequency content providing most information. Such data could be sampled efficiently with a non-uniform schedule that increases the sampling rate only during activity bursts.
METHODS: A real time and adaptive algorithm is proposed to select the sampling rate, in order to reduce the number of measured samples, but still recording the main information. The algorithm is based on a neural network which predicts the subsequent samples and their uncertainties, requiring a measurement only when the risk of the prediction is larger than a selectable threshold.
RESULTS: Four examples of application to biomedical data are discussed: electromyogram, electrocardiogram, electroencephalogram, and body acceleration. Sampling rates are reduced under the Nyquist limit, still preserving an accurate representation of the data and of their power spectral densities (PSD). For example, sampling at 60% of the Nyquist frequency, the percentage average rectified errors in estimating the signals are on the order of 10% and the PSD is fairly represented, until the highest frequencies. The method outperforms both uniform sampling and compressive sensing applied to the same data.
CONCLUSION: The discussed method allows to go beyond Nyquist limit, still preserving the information content of non-stationary biomedical signals. It could find applications in body sensor networks to lower the number of wireless communications (saving sensor power) and to reduce the occupation of memory.
Copyright © 2016 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Accelerometer; Compressive sensing; Electrocardiography (ECG); Electroencephalography (EEG); Electromyography (EMG); Non-uniform sampling; Nyquist limit

Mesh:

Year:  2016        PMID: 26917540     DOI: 10.1016/j.compbiomed.2016.02.004

Source DB:  PubMed          Journal:  Comput Biol Med        ISSN: 0010-4825            Impact factor:   4.589


  3 in total

Review 1.  Wearable biosensors for healthcare monitoring.

Authors:  Jayoung Kim; Alan S Campbell; Berta Esteban-Fernández de Ávila; Joseph Wang
Journal:  Nat Biotechnol       Date:  2019-02-25       Impact factor: 54.908

2.  Hybrid Prediction Method for ECG Signals Based on VMD, PSR, and RBF Neural Network.

Authors:  Fuying Huang; Tuanfa Qin; Limei Wang; Haibin Wan
Journal:  Biomed Res Int       Date:  2021-03-15       Impact factor: 3.411

3.  DynDSE: Automated Multi-Objective Design Space Exploration for Context-Adaptive Wearable IoT Edge Devices.

Authors:  Giovanni Schiboni; Juan Carlos Suarez; Rui Zhang; Oliver Amft
Journal:  Sensors (Basel)       Date:  2020-10-27       Impact factor: 3.847

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.