| Literature DB >> 26909703 |
Nafisur Rahman1, Sumaiya Khan2.
Abstract
Ampicillin exhibited a negative and a positive cotton effects on the circular dichroism (CD) spectra in the wavelength range of 200-280nm. Cloxacillin showed a positive cotton band peaking at 228nm. Three sensitive, precise and accurate CD spectroscopic methods have been developed for the determination of ampicillin and cloxacillin. Method A was used for the determination of ampicillin in presence of cloxacillin by measuring ellipticity at 206nm. Method B and C were employed to determine ampicillin and cloxacillin based on evaluation of ellipticity at 233nm and 228nm, respectively. Methods A, B and C showed linearity in the concentration range of 10-40μgmL(-1), 5-40μgmL(-1) ampicillin and 10-80μgmL(-1) cloxacillin, respectively. The method A was successfully applied to the determination of ampicillin in commercial dosage forms containing equivalent amount of cloxacillin.Entities:
Keywords: Ampicillin; Circular dichroism spectra; Cloxacillin; Pharmaceutical formulation; Validation
Mesh:
Substances:
Year: 2016 PMID: 26909703 DOI: 10.1016/j.saa.2016.02.009