Literature DB >> 26902984

A new robotic assistance system for percutaneous CT-guided punctures: Initial experience.

Simon Groetz1, Kai Wilhelm2, Winfried Willinek3, Claus Pieper1, Hans Schild1, Daniel Thomas1.   

Abstract

PURPOSE: To evaluate the feasibility of CT-guided bone biopsies using a novel robotic needle guide.
MATERIAL AND METHODS: The robotic needle guide iSYS 1.3 (iSYS Medzintechnik, GmbH, Kitzbuehel, Austria) mounted on the standard table of a CT scanner was used for all studies. For preclinical testing, eight vertebral bodies of dead swine were biopsied, trying to place the needle in the center of the vertebral body via a transpedicular access. For clinical evaluation, bone biopsies were taken in three different patients with ambiguous bone lesions. All biopsies were performed under general anesthesia, using a 12G bone biopsy needle.
RESULTS: The animal studies demonstrated that the biopsy needle could be placed accurately in the center of the vertebral body in all cases. No readjustment was necessary, the CT scans demonstrated an intrapedicular trajectory avoiding the spinal canal or the neural foramina. Subsequently, following the animal studies, all biopsies could be performed successfully. Needle placement was accurate without any need for readjustment. No complications occurred during the procedure.
CONCLUSION: Using the iSYS 1.3 allows for accurate and simple stereotactic biopsies of bone lesions, avoiding the need for needle readjustment. The systems may offer even less experienced teams to take biopsies in regions which are difficult to access.

Entities:  

Keywords:  CT-guided; biopsy; image guidance; navigation; puncture; robot

Mesh:

Year:  2015        PMID: 26902984     DOI: 10.3109/13645706.2015.1110825

Source DB:  PubMed          Journal:  Minim Invasive Ther Allied Technol        ISSN: 1364-5706            Impact factor:   2.442


  3 in total

1.  Robotic needle insertion during computed tomography fluoroscopy-guided biopsy: prospective first-in-human feasibility trial.

Authors:  Takao Hiraki; Tetsushi Kamegawa; Takayuki Matsuno; Jun Sakurai; Toshiyuki Komaki; Takuya Yamaguchi; Koji Tomita; Mayu Uka; Yusuke Matsui; Toshihiro Iguchi; Hideo Gobara; Susumu Kanazawa
Journal:  Eur Radiol       Date:  2019-08-23       Impact factor: 5.315

2.  Puncture accuracy of an optical tracked robotic aiming device-a phantom study.

Authors:  Yannick Scharll; Sofia Letrari; Gregor Laimer; Peter Schullian; Reto Bale
Journal:  Eur Radiol       Date:  2022-06-09       Impact factor: 7.034

3.  Infrared needle mapping to assist biopsy procedures and training.

Authors:  Bruce Shar; John Leis; John Coucher
Journal:  Healthc Technol Lett       Date:  2018-01-26
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.