Henrik Casimir Ahn1, Jacek Baranowski2, Lars-Goran Dahlin3, Niels Erik Nielsen4. 1. Department of Cardiothoracic Surgery, Institution of Medical and Health Sciences, Linkoping University, Linkoping, Sweden. Electronic address: henrik.casimir.ahn@liu.se. 2. Department of Physiology, Institution of Medical and Health Sciences, Linkoping University, Linkoping, Sweden. 3. Department of Cardiothoracic Surgery, Institution of Medical and Health Sciences, Linkoping University, Linkoping, Sweden. 4. Department of Cardiology, Institution of Medical and Health Sciences, Linkoping University, Linkoping, Sweden.
Abstract
BACKGROUND: The purpose of this study was to report the use of a transvenous transseptal approach using a stent valve in patients with degenerated biological mitral valve prostheses, regurgitation after mitral repair, and native mitral stenosis. METHODS: Ten patients (median age, 74 years; range, 20-89 years; 5 men and 5 women) with degenerated mitral bioprosthetic valves (n = 7), failed mitral repair (n = 1), or calcified native stenotic valves (n = 2) underwent transvenous implantation of a stent valve. RESULTS: The procedure was initially successful in all patients. Predilation was performed for balloon sizing only in the 2 patients with native mitral stenosis. The stent valve was deployed during 1 period of rapid pacing. A guidewire, as a loop from the right femoral vein and through the left ventricular apex, facilitated a good angle and secure positioning of the stent valve. An ultrasonographically guided puncture of the apex was carried out in 6 patients, and in the other 4 we performed a minithoracotomy before apical puncture. All valves were implanted in a good position with improved function and without significant paravalvular leakage (PVL). There were no periprocedural deaths. The 30-day survival was 80% (8 of 10 patients), and 60% (6 of 10) of patients were still alive a median time of 290 days after the procedure. CONCLUSIONS: Transvenous transseptal implantation of a stent valve was performed in 10 patients with mitral valve disease, with good early functional results. These high-risk patients must be carefully selected by a multidisciplinary team because the procedure carries a high mortality.
BACKGROUND: The purpose of this study was to report the use of a transvenous transseptal approach using a stent valve in patients with degenerated biological mitral valve prostheses, regurgitation after mitral repair, and native mitral stenosis. METHODS: Ten patients (median age, 74 years; range, 20-89 years; 5 men and 5 women) with degenerated mitral bioprosthetic valves (n = 7), failed mitral repair (n = 1), or calcified native stenotic valves (n = 2) underwent transvenous implantation of a stent valve. RESULTS: The procedure was initially successful in all patients. Predilation was performed for balloon sizing only in the 2 patients with native mitral stenosis. The stent valve was deployed during 1 period of rapid pacing. A guidewire, as a loop from the right femoral vein and through the left ventricular apex, facilitated a good angle and secure positioning of the stent valve. An ultrasonographically guided puncture of the apex was carried out in 6 patients, and in the other 4 we performed a minithoracotomy before apical puncture. All valves were implanted in a good position with improved function and without significant paravalvular leakage (PVL). There were no periprocedural deaths. The 30-day survival was 80% (8 of 10 patients), and 60% (6 of 10) of patients were still alive a median time of 290 days after the procedure. CONCLUSIONS: Transvenous transseptal implantation of a stent valve was performed in 10 patients with mitral valve disease, with good early functional results. These high-risk patients must be carefully selected by a multidisciplinary team because the procedure carries a high mortality.