Comparison of long-labeled pseudo-continuous arterial spin labeling (ASL) features between young and elderly adults: special reference to parameter selection.

Yasuhiro Fujiwara1, Tsuyoshi Matsuda2, Masayuki Kanamoto3, Tatsuro Tsuchida4, Kazunobu Tsuji4, Nobuyuki Kosaka4, Toshiki Adachi3, Hirohiko Kimura5.   

Abstract

BACKGROUND: The signal intensity obtained by arterial spin labeling (ASL) depends not only on perfusion signal, but also on arterial transit time (ATT). Although ATT has a more significant effect on accurate regional cerebral blood flow (CBF) calculations, the multiple post-labeling delay (PLD) approach is difficult to use in routine examinations.
PURPOSE: To optimize imaging parameters for labeling duration (LD) and PLD and to confirm their validity in long-labeled pseudo-continuous ASL.
MATERIAL AND METHODS: The perfusion signal was simulated in four LDs and theoretical signal-to-noise ratio efficiency (SNReff) was calculated. In vivo studies were performed on a 3.0 T magnetic resonance imaging (MRI) scanner and 15 volunteers were categorized into either the young or elderly adult groups. We compared the differences in CBF values with or without ATT correction.
RESULTS: Regarding signal simulation, perfusion signal increased with the length of LD. SNReff also improved with LD, but SNReff plateaued at an LD of 3.0 s. As for the in vivo study, SNR linearly increased along with the LD. The CBF differences with the correction of ATT were larger in the elderly adult group. This trend was most prominent in the longer ATT area in the occipital cortical region.
CONCLUSION: A combination of imaging settings of LD = 3.5 s and PLD = 2.0 s were suggested as optimal imaging parameters for allowing acceptable CBF quantification and sufficient SNR in both young and elderly individuals. © The Foundation Acta Radiologica 2016.

Entities:  

Keywords:  Magnetic resonance imaging (MRI); arterial spin labeling (ASL); arterial transit time (ATT); cerebral blood flow; perfusion

Mesh:

Substances:

Year:  2016        PMID: 26893213     DOI: 10.1177/0284185116632387

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  1 in total

1.  Optimization of post-label delay in single-phase arterial spin labeling (ASL) using multi-phase ASL in four-dimensional magnetic resonance angiography.

Authors:  Tsuyoshi Oshita; Hidetake Yabuuchi; Masanobu Osame
Journal:  Neuroradiol J       Date:  2019-04-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.