Literature DB >> 26889881

First-in-Man Transcervical Surgical Aortic Valve Replacement Using the CoreVista System.

Otto E Dapunt1, Olev Luha, Adrian Ebner, Piotr Sonecki, Cristiano Spadaccio, Fraser W H Sutherland.   

Abstract

OBJECTIVE: This study aimed to evaluate a novel device system for surgical aortic valve replacement (SAVR) using a unique new less invasive access approach. The hypothesis is that SAVR can be performed through a short transverse incision in the neck, similar to that used for transcervical thymectomy avoiding chest disruption.
METHODS: A new device system was developed to provide retraction, step-by-step illumination, and on-screen visualization for the new approach. Preliminary feasibility studies were performed in cadavers. Comprehensive risk analysis was performed, and training was implemented in Thiel preserved cadavers. For the first-in-man clinical case, a 63-year-old woman with symptomatic critical aortic stenosis (The Society of Thoracic Surgeons risk, 11%) and heavily calcified aortic valve was selected. A short transverse incision was made in the neck; the device was introduced, and the sternum was elevated; femorofemoral cardiopulmonary bypass was established; substernal dissection was guided by the sequenced illumination, and high-definition visualization was provided by the device, allowing for optimal exposition of the aorta and aortic valve; and a 23-mm Medtronic ENABLE sutureless valve prosthesis was implanted. Procedure success was evaluated according to the standardized composite end point definition of "device success" proposed by the Valve Academic Research Consortium.
RESULTS: Access, delivery, and deployment of the valve prosthesis were successful. The correct position and intended performance of the valve were demonstrated (mean gradient, 6 mm Hg; aortic valve area, 2.5 cm) with the absence of moderate or severe prosthetic aortic regurgitation. Only one valve prosthesis was used.
CONCLUSIONS: Transcervical SAVR with sutureless valve is feasible using this novel access system. The new approach has potential to offer patients substantially shorter stay and fewer, less serious complications, as has been observed in transcervical thymectomy. Further studies are merited.

Entities:  

Mesh:

Year:  2016        PMID: 26889881     DOI: 10.1097/IMI.0000000000000228

Source DB:  PubMed          Journal:  Innovations (Phila)        ISSN: 1556-9845


  2 in total

Review 1.  Minimally invasive aortic valve surgery.

Authors:  Lorenzo Di Bacco; Antonio Miceli; Mattia Glauber
Journal:  J Thorac Dis       Date:  2021-03       Impact factor: 2.895

2.  Transcervical approach for left ventricular assist device outflow graft: A proof-of-concept implantation study.

Authors:  Fraser William Havern Sutherland; George Gradinariu; Philip Curry
Journal:  JTCVS Tech       Date:  2021-01-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.