Woojun Jung1, Minho Hwang1, Young Ju Won1, Byung Gun Lim1, Myoung-Hoon Kong1, Il-Ok Lee1.
Abstract
BACKGROUND: Electromyography and acceleromyography are common neuromuscular monitoring devices. However, questions still remain regarding the use of acceleromyography in children. This study compared the calibration success rates and intubation conditions in children after obtaining the maximal blockade depending on each of the devices.Entities:
Keywords: Acceleromyography; Electromyography; Neuromuscular monitoring; Preschool child; Rocuronium
Year: 2016 PMID: 26885297 PMCID: PMC4754261 DOI: 10.4097/kjae.2016.69.1.21
Source DB: PubMed Journal: Korean J Anesthesiol ISSN: 2005-6419
Assessment of Intubating Conditions
| Variables | Intubating conditions | ||
|---|---|---|---|
| Acceptable | Unacceptable Poor | ||
| Excellent | Good | ||
| Ease of laryngoscopy (jaw relaxation) | Easy | Fair | Difficult |
| Vocal cord position | Abducted | Intermediate | Closed |
| Vocal cord movement | None | Moving | Closing |
| Airway reaction (coughing) | None | Diaphragm | Sustained (> 10 sec) |
| Movement of the limbs | None | Slight | Vigorous |
Intubating conditions; Excellent: all criteria are 'excellent', Good: all criteria are either 'excellent' or 'good', Poor: the presence of a single criterion listed under 'poor.'
Calibration Success Rate and Time
| TOF Watch SX group | NMT group | P value | |
|---|---|---|---|
| Success rate of 1st calibration (%) | 46.9% (23/49) | 64.5% (31/48) | > 0.8 |
| Success rate of 2nd calibration (%) | 61.2% (30/49) | 79.1% (38/48) | > 0.8 |
| Calibration time (sec) | 28.1 ± 13.4 | 16.7 ± 11.0 | 0.012 |
Data are number of patients or mean ± SD. The average of the first and second success rate was similar in both groups (P value > 0.8), but calibration time of the group N (16.7 sec) was significantly shorter than group T (28.1 sec) (P value = 0.012). TOF Watch SX group: acceleromyography, Group T: NMT group, Electromyography: Group N.
Causes of Calibration Failure
| Cause of failure | TOF Watch SX group | NMT group | ||
|---|---|---|---|---|
| 1st calibration | 2nd calibration | 1st calibration | 2nd calibration | |
| High resistance | 4 | 4 | 12 | 10 |
| Unstable transducer | 16 | 11 | 0 | 0 |
| Involuntary movement | 6 | 4 | 5 | 0 |
TOF Watch SX group: acceleromyography, Group T: NMT group, Electromyography: group N.
Comparison of BIS, Onset Time of Rocuronium and Intubation Condition
| TOF Watch SX group | NMT group | P value | |
|---|---|---|---|
| BIS at | |||
| Baseline | 81.6 ± 10.0 | 85.9 ± 9.9 | 0.066 |
| Calibration | 50.8 ± 7.7 | 54.6 ± 10.1 | 0.072 |
| Rocuronium injection | 53.1 ± 9.0 | 55.4 ± 10.7 | 0.356 |
| Intubation | 57.9 ± 5.4 | 62.1 ± 7.7 | 0.017 |
| After intubation | 60.1 ± 5.6 | 65.1 ± 6.5 | 0.003 |
| Onset time (sec) | 63.9 ± 18.8 | 73.6 ± 18.9 | 0.042 |
| Intubation condition | 1.86 ± 0.50 | 2.27 ± 0.65 | 0.007 |
The rocuronium onset time was measured by seconds from the intravenous administration of rocuronium to the twitch height suppressed to 0. The classification of the intubation conditions, as excellent, good, and poor, following the criteria shown in Table 1 were measured as scores 3, 2, and 1, respectively, and statistical calculations were performed. TOF Watch SX group: acceleromyography, Group T: NMT group, Electromyography: group N.