A Comparison of Three Widely Used Immunoassay Systems in Cortisol Measurement.

Nihal Yücel, Özlem Çakir Madenci, Aycan Bölük, Lale Köroğlu, Yusuf Temel, Asuman Orçun.   

Abstract

BACKGROUND: Interassay variability is one of the challenging issues of routine clinical laboratory practice. Commercial plasma cortisol immunoassays are also subject to this issue. In this study, we intended to evaluate the interchangeability of cortisol results of three widely used immunoassay systems.
METHODS: The cortisol values of 150 serum samples measured by three immunoassay systems, Beckman Coulter DXI 800, Roche Modular E170, and Siemens Immulite 2500, were compared.
RESULTS: A degree of proportional biases was observed between all three methods; DXI 800 showed the worst biases with the other two systems (slope values 0.67 and 0.77 with E170 and Immulite 2500, respectively). DXI 800 showed poor agreement with other methods (CCC: 0.83 and 0.87, respectively). There was a moderate agreement between E170 and Immulite 2500 (CCC: 0.92).
CONCLUSIONS: All three methods showed a degree of variability among themselves. DXI 800 results were not interchangeable with the other two systems.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26882820

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  1 in total

1.  Long-acting Porcine Sequence ACTH (Acton Prolongatum) Stimulation Test is a Reliable Alternative Test as Compared to the Gold Standard Insulin Tolerance Test for the Diagnosis of Adrenal Insufficiency.

Authors:  Sridevi Atluri; Vijaya Sarathi; Amit Goel; Shivaprasad Channabasappa; Shailaja Alapaty; Melkunte S Dhananjaya; Ramdas Barure; Gautam Kolla
Journal:  Indian J Endocrinol Metab       Date:  2022-04-27
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.