Literature DB >> 26872735

Aortic Valve Replacement With Carpentier-Edwards: Hemodynamic Outcomes for the 19-mm Valve.

Ji Hoon You1, Dong Seop Jeong2, Kiick Sung2, Wook Sung Kim2, K C Carriere3, Young Tak Lee2, Pyo Won Park4.   

Abstract

BACKGROUND: To compare hemodynamic performance and clinical outcomes after aortic valve replacement for aortic stenosis with the 19-mm Carpentier-Edwards pericardial bioprosthesis versus larger valves.
METHODS: Between January 1998 and December 2013, 447 consecutive patients underwent aortic valve replacement for aortic stenosis with the Carpentier-Edwards Perimount (n = 61) or Magna bioprostheses (n = 386). Based on the implanted valve size, the patients were classified into three groups: a 19-mm group (n = 54), a 21-mm group (n = 154), and a 23-mm to 27-mm group (n = 239). The in vivo effective orifice area index was measured by transthoracic echocardiography 12 months after operation (n = 331). The mean follow-up time was 4.9 ± 3.5 (maximum 15.4) years.
RESULTS: There were three early deaths (0.7%). At 10 years, overall survival (84.1%) was unaffected by patient-prosthesis mismatch (18.7%, 62 patients), and freedom from structural valve deterioration and endocarditis was 100% and 97.1%, respectively. Although the 19-mm group was significantly older and had a higher incidence of patient-prosthesis mismatch (n = 14, 30.4%), there were no significant differences in early outcomes, overall survival, cardiac-related mortality, or serial reduction of left ventricular mass index in comparison with patients with a larger bioprostheses. Independent risk factors for all-cause mortality were age, male gender, combined coronary artery bypass graft, and low hemoglobin level.
CONCLUSION: The Carpentier-Edwards pericardial bioprosthesis appears to be associated with acceptable clinical outcomes and hemodynamic profile.
Copyright © 2016 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 26872735     DOI: 10.1016/j.athoracsur.2015.11.016

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  3 in total

1.  Late clinical outcomes of aortic valve replacement with Carpentier-Edwards pericardial valves.

Authors:  Hyoung Woo Chang; Wook Sung Kim; Joong Hyun Ahn; Keumhee C Carriere; Dong Seop Jeong; Yang Hyun Cho; Kiick Sung; Pyo Won Park
Journal:  J Thorac Dis       Date:  2019-12       Impact factor: 2.895

2.  Late clinical outcomes after mechanical aortic valve replacement for aortic stenosis: old versus new prostheses.

Authors:  Heemoon Lee; Kiick Sung; Wook Sung Kim; Dong Seop Jeong; Joong Hyun Ahn; Keumhee Chough Carriere; Pyo Won Park
Journal:  J Thorac Dis       Date:  2018-06       Impact factor: 2.895

3.  Long-Term Results (up to 20 Years) of 19 mm or Smaller Prostheses in the Aortic Position. Does Size Matter? A Propensity-Matched Survival Analysis.

Authors:  Horea Feier; Andrei Grigorescu; Lucian Falnita; Oana Rachita; Marian Gaspar; Constantin T Luca
Journal:  J Clin Med       Date:  2021-05-11       Impact factor: 4.241

  3 in total