| Literature DB >> 26867552 |
Myong Ki Baeg1, Sang-Woo Kim1, Sun-Hye Ko2, Yoon Bum Lee1, Seawon Hwang2, Bong-Woo Lee2, Hye Jin Choi2, Jae Myung Park1, In-Seok Lee1, Yong-Seog Oh3, Myung-Gyu Choi1.
Abstract
BACKGROUND/AIMS: Patients with cardiac implantable electronic devices (CIEDs) undergoing endoscopic electrosurgery (EE) are at a risk of electromagnetic interference (EMI). We aimed to analyze the effects of EE in CIED patients.Entities:
Keywords: Defibrillators, implantable; Electromagnetic interference; Endoscopic electrosurgery; Permanent pacemaker
Year: 2016 PMID: 26867552 PMCID: PMC4821527 DOI: 10.5946/ce.2015.023
Source DB: PubMed Journal: Clin Endosc ISSN: 2234-2400
Demographic Characteristics of Patients with Cardiac Implanted Electronic Devices Undergoing Endoscopic Electrosurgery
| Characteristic | Total | Pacemaker | ICD |
|---|---|---|---|
| No. of patients | 49 | 43 | 6 |
| No. of procedures | 59 | 50 | 9 |
| Age, yr | 69.1±9.1 | 71.1±8.2 | 57.7±4.4 |
| Male sex | 39 | 30 | 6 |
| Reason for device implantation | |||
| Sick sinus syndrome | 32 | 32 | 0 |
| Complete atrioventricular block | 8 | 8 | 0 |
| Second degree atrioventricular block | 5 | 5 | 0 |
| Atrial fibrillation with slow ventricular response | 2 | 2 | 0 |
| Junctional bradycardia | 2 | 2 | 0 |
| Idiopathic ventricular tachycardia | 8 | 0 | 8 |
| Brugada syndrome | 1 | 0 | 1 |
| Unknown | 1 | 1 | 0 |
| Time from device implantation to endoscopic surgery, day | 1,844±1,640 | 2,006±1,722 | 982±653 |
| Type of endoscopic electrosurgery | |||
| Colon snare polypectomy | 44 | 35 | 9 |
| Colon ESD | 1 | 1 | 0 |
| Gastric snare polypectomy | 1 | 1 | 0 |
| Gastric ESD | 5 | 5 | 0 |
| ERCP with EST | 8 | 8 | 0 |
| Admission status | |||
| Admission | 26 | 26 | 0 |
| Outpatient | 33 | 24 | 9 |
Values are presented as mean±SD.
ICD, implantable cardioverter-defibrillator; ESD, endoscopic submucosal dissection; ERCP, endoscopic retrograde cholangiopancreatography; EST, endoscopic sphincterotomy.
Device Characteristics of Patients with Cardiac Implanted Electronic Devices Undergoing Endoscopic Electrosurgery
| Characteristic | Total | Pacemaker | ICD |
|---|---|---|---|
| Device type | |||
| AAI | 50 | 50 | 0 |
| DDD | 3 | 0 | 3 |
| DDDR | 1 | 0 | 1 |
| VVI | 5 | 0 | 5 |
| Device manufacturer | |||
| Boston Scientific | 8 | 1 | 7 |
| St. Jude | 4 | 2 | 2 |
| Biotronik | 12 | 12 | 0 |
| Medtronic | 20 | 20 | 0 |
| ELA Medical | 4 | 4 | 0 |
| Vitatron | 9 | 9 | 0 |
| Unknown | 2 | 2 | 0 |
ICD, implantable cardioverter-defibrillator; AAI, atrial pacing, atrial sensing, inhibited by atrial event; DDD, dual chamber pacing, dual chamber sensing, dual function; DDDR, dual chamber pacing, dual chamber sensing, dual function, rate responsive pacing; VVI, ventricle pacing, ventricle sensing, inhibited by ventricle event.
Follow-Up Data of Patients with Cardiac Implanted Electronic Devices Undergoing Endoscopic Electrosurgery
| Variable | Total | Pacemaker | ICD |
|---|---|---|---|
| Time to cardiology outpatient clinic visit, day | 46±47 (53/59[ | 44±46 (44) | 59±54 (9) |
| Time to initial electrical follow-up, day | 111±119 (55/59[ | 103±122 (46) | 154±92 (9) |
| Type of initial electrical follow-up[ | |||
| Electrocardiogram | 31 | 28 | 3 |
| 24-Hour Holter monitoring | 9 | 6 | 3 |
| Pacemaker interrogation | 23 | 17 | 6 |
| Not done | 5 | 5 | 0 |
| Time to next pacemaker interrogation after endoscopic electrosurgery | 226±222 (55) | 242±240 (46) | 173±113 (9) |
| Postprocedure changes noted on electrical follow-up | 2 | 2 | 0 |
Values are presented as mean±SD (number). Differences between outpatient visit and electrical follow-ups are due to some electrical follow-ups being performed by departments other than cardiology.
ICD, implantable cardioverter-defibrillator.
Two follow-up losses, two followed up by the oncology department for underlying malignancies, one unable to visit the hospital as the patient was staying at a facility for the disabled and one whose follow-up date had not come up;
Two follow-up losses, one who visited the cardiology outpatient clinic but was deemed asymptomatic and scheduled for regular pacemaker interrogation at a later date and one who has not yet visited the cardiology outpatient clinic;
Some electrical follow-up types were carried out simultaneously.
Pacemaker Interrogation Results of Devices with Recordings during Endoscopic Electrosurgery
| Variable | Total | Pacemaker | ICD |
|---|---|---|---|
| No. of devices with recordings | 31 | 22 | 9 |
| No events recorded | 29 | 20 | 9 |
| Events recorded | 2 | 2 | 0 |
ICD, implantable cardioverter-defibrillator.