Motion correction and frequency stabilization for MRS of the human spinal cord.

Andreas Hock1,2, Anke Henning1,3.   

Abstract

Subject motion is challenging for MRS, because it can falsify results. For spinal cord MRS in particular, subject movement is critical, since even a small movement > 1 mm) can lead to a voxel shift out of the desired measurement region. Therefore, the identification of motion corrupted MRS scans is essential. In this investigation, MR navigators acquired simultaneously with the MRS data are used to identify a displacement of the spinal cord due to subject motion. It is shown that navigators are able to recognize substantial subject motion (>1 mm) without impairing the MRS measurement. In addition, navigators are easy to apply to the measurement, because no additional hardware and just a minor additional user effort are needed. Moreover, no additional scan time is required, because navigators can be applied in the deadtime of the MRS sequence. Furthermore, in this work, retrospective motion correction combined with frequency stabilization is presented by combining navigators with non-water-suppressed (1)H-MRS, resulting in an improved spectral quality of the spinal cord measurements.
Copyright © 2016 John Wiley & Sons, Ltd.

Entities:  

Keywords:  MRS; human spinal cord; motion detection; navigator

Mesh:

Substances:

Year:  2016        PMID: 26867133     DOI: 10.1002/nbm.3487

Source DB:  PubMed          Journal:  NMR Biomed        ISSN: 0952-3480            Impact factor:   4.044


  1 in total

1.  Motion correction methods for MRS: experts' consensus recommendations.

Authors:  Ovidiu C Andronesi; Pallab K Bhattacharyya; Wolfgang Bogner; In-Young Choi; Aaron T Hess; Phil Lee; Ernesta M Meintjes; M Dylan Tisdall; Maxim Zaitzev; André van der Kouwe
Journal:  NMR Biomed       Date:  2020-07-20       Impact factor: 4.044

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.