A Parametric Computational Study of the Impact of Non-circular Configurations on Bioprosthetic Heart Valve Leaflet Deformations and Stresses: Possible Implications for Transcatheter Heart Valves.

Nandini Duraiswamy1, Jason D Weaver2, Yasamin Ekrami2, Stephen M Retta2, Changfu Wu3.   

Abstract

Although generally manufactured as circular devices with symmetric leaflets, transcatheter heart valves can become non-circular post-implantation, the impact of which on the long-term durability of the device is unclear. We investigated the effects of five non-circular (EllipMajor, EllipMinor, D-Shape, TriVertex, TriSides) annular configurations on valve leaflet stresses and valve leaflet deformations through finite element analysis. The highest in-plane principal stresses and strains were observed under an elliptical configuration with an aspect ratio of 1.25 where one of the commissures was on the minor axis of the ellipse. In this elliptical configuration (EllipMinor), the maximum principal stress increased 218% and the maximum principal strain increased 80% as compared with those in the circular configuration, and occurred along the free edge of the leaflet whose commissures were not on the minor axis (i.e., the "stretched" leaflet). The D-Shape configuration was similar to this elliptical configuration, with the degree to which the leaflets were stretched or sagging being less than the EllipMinor configuration. The TriVertex and TriSides configurations had similar leaflet deformation patterns in all three leaflets and similar to the Circular configuration. In the D-Shape, TriVertex, and TriSides configurations, the maximum principal stress was located near the commissures similar to the Circular configuration. In the EllipMinor and EllipMajor configurations, the maximum principal stress occurred near the center of the free edge of the "stretched" leaflets. These results further affirm recommendations by the International Standards Organization (ISO) that pre-clinical testing should consider non-circular configurations for transcatheter valve durability testing.

Entities:  

Keywords:  FEA; Leaflet durability; Non-circular deformation; TAVR; THV; Transcatheter heart valve

Mesh:

Year:  2016        PMID: 26864541     DOI: 10.1007/s13239-016-0259-9

Source DB:  PubMed          Journal:  Cardiovasc Eng Technol        ISSN: 1869-408X            Impact factor:   2.495


  3 in total

Review 1.  Principles of TAVR valve design, modelling, and testing.

Authors:  Oren M Rotman; Matteo Bianchi; Ram P Ghosh; Brandon Kovarovic; Danny Bluestein
Journal:  Expert Rev Med Devices       Date:  2018-10-29       Impact factor: 3.166

2.  Fluid-Structure Interaction Models of Bioprosthetic Heart Valve Dynamics in an Experimental Pulse Duplicator.

Authors:  Jae H Lee; Alex D Rygg; Ebrahim M Kolahdouz; Simone Rossi; Stephen M Retta; Nandini Duraiswamy; Lawrence N Scotten; Brent A Craven; Boyce E Griffith
Journal:  Ann Biomed Eng       Date:  2020-02-07       Impact factor: 3.934

3.  Advancing Regulatory Science With Computational Modeling for Medical Devices at the FDA's Office of Science and Engineering Laboratories.

Authors:  Tina M Morrison; Pras Pathmanathan; Mariam Adwan; Edward Margerrison
Journal:  Front Med (Lausanne)       Date:  2018-09-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.