Joel A Ramirez1, Patrick T O'Shaughnessy1.
Abstract
The objective of this study was to determine the effect of several simulated air environmental conditions on the particle penetration and the breathing resistance of two N95 filtering facepiece respirator (FFR) models. The particle penetration and breathing resistance of the respirators were evaluated in a test system developed to mimic inhalation and exhalation breathing while relative humidity and temperature were modified. Breathing resistance was measured over 120 min using a calibrated pressure transducer under four different temperature and relative humidity conditions without aerosol loading. Particle penetration was evaluated before and after the breathing resistance test at room conditions using a sodium chloride aerosol measured with a scanning mobility particle sizer. Results demonstrated that increasing relative humidity and lowering external temperature caused significant increases in breathing resistance (p < 0.001). However, these same conditions did not influence the penetration or most penetrating particle size of the tested FFRs. The increase in breathing resistance varied by FFR model suggesting that some FFR media are less influenced by high relative humidity.Entities:
Keywords: Breathing resistance; cyclic flow; filtering facepiece respirators; particle penetration
Mesh:
Substances:
Year: 2016 PMID: 26861653 DOI: 10.1080/15459624.2016.1143950
Source DB: PubMed Journal: J Occup Environ Hyg ISSN: 1545-9624 Impact factor: 2.155