Correlation between scatter radiation dose at height of operator's eye and dose to patient for different angiographic projections.

Fernando Leyton1, Maria S Nogueira2, Luiz A Gubolino3, Makyson R Pivetta4, Carlos Ubeda5.   

Abstract

Studies have reported cases of radiation-induced cataract among cardiology professionals. In view of the evidence of epidemiological studies, the ICRP recommends a new threshold for opacities and a new radiation dose to eye lens limit of 20mSv per year for occupational exposure. The aim of this paper is to report scattered radiation doses at the height of the operator's eye in an interventional cardiology facility without considering radiation protection devices and to correlate these values with different angiographic projections and operational modes. Measurements were taken in a cardiac laboratory with an angiography X-ray system equipped with flat-panel detector. PMMA plates of 30×30×5cm were used with a thickness of 20cm. Measurements were taken in two fluoroscopy modes (low and normal, 15pulses/s) and in cine mode (15frames/s). Four angiographic projections were used: anterior posterior; lateral; left anterior oblique caudal (spider); and left anterior oblique cranial, with a cardiac protocol for patients weighing between 70 and 90kg. Measurements of phantom entrance dose rate and scatter dose rate were performed with two Unfors Xi plus detectors. The detector measuring scatter radiation was positioned at the usual distance of the cardiologist's eyes during working conditions. There is a good linear correlation between the kerma area product and scatter dose at the lens. Experimental correlation factors of 2.3, 12.0, 12.2 and 17.6μSv/Gycm2 were found for different projections. PMMA entrance dose rates for low and medium fluoroscopy and cine modes were 13, 39 and 282mGy/min, respectively, for AP projection.
Copyright © 2016 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Ambient dose equivalent; Angiography; Eye lens dose; Occupational dose; Patient dose; Scatter radiation dose

Mesh:

Year:  2016        PMID: 26857629     DOI: 10.1016/j.apradiso.2016.01.013

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  4 in total

1.  Beam projections and radiation exposure in transradial and transfemoral approaches during coronary angiography.

Authors:  Ali Tarighatnia; Amirhossien Mohammadalian; Morteza Ghojazade; Leili Pourafkari; Alireza Farajollahi
Journal:  Anatol J Cardiol       Date:  2017-08-02       Impact factor: 1.596

2.  Taller staff occupationally exposed to less radiation to the temple in cardiac procedures, but risk higher doses during vascular cases.

Authors:  Kelly S Wilson-Stewart; Davide Fontanarosa; Dan Li; Chris C Drovandi; Rebecca K Anderson; Jamie V Trapp
Journal:  Sci Rep       Date:  2020-09-30       Impact factor: 4.379

3.  Occupational and Patient Radiation Dose and Quality Implications of Femoral Access Imaging During Coronary Angiography.

Authors:  Kelly Wilson-Stewart; Davide Fontanarosa; Eva Malacova; Steven Gett; Allan Kruger; Jamie V Trapp
Journal:  J Multidiscip Healthc       Date:  2021-07-12

4.  Assessment of eye doses to staff involved in interventional cardiology procedures in Kuwait.

Authors:  Meshari Alnaaimi; Mousa Alduaij; Faisal Shenawy; Musab Algaily; Talal Mohammedzein; Mohamed Shaaban; Madan M Rehani
Journal:  Radiat Environ Biophys       Date:  2021-08-05       Impact factor: 1.925

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.