Literature DB >> 26857191

Low risk of electromagnetic interference between smartphones and contemporary implantable cardioverter defibrillators.

Haran Burri1, Louis Paulin Mondouagne Engkolo2, Nicolas Dayal2, Abdul Etemadi2, Anne-Marie Makhlouf2, Carine Stettler2, Florence Trentaz2.   

Abstract

AIMS: Manufacturers of implantable cardioverter defibrillators (ICDs) recommend that cell phones be maintained at a distance of ∼15 cm from the implanted device in order to avoid risk of dysfunction due to electromagnetic interference (EMI). Data relating to this issue are outdated and do not reflect modern technology. Our aim was to evaluate whether EMI is still an issue with contemporary ICDs and smartphones. METHODS AND
RESULTS: Consecutive patients implanted with a wireless-enabled ICD were tested for potential interference with two models of recent 4G smartphones in conditions intended to maximize risk of EMI. A magnet effect (due to the phone speakers) was tested by placing the smartphones in the standby mode directly over the ICD generator. The presence of EMI artefacts on the real-time electrograms was evaluated by placing the smartphones in the standby, dialling, and operating modes directly over the generator casing and over the parasternal region in the vicinity of the ventricular lead. A total of 63 patients equipped with 29 different models of single, dual, or biventricular ICDs from five major manufacturers were included. None of the patients showed any evidence of interference with the smartphones during any of the 882 tests.
CONCLUSION: The risk of EMI between modern smartphones and contemporary ICDs is low. This is probably due to the filters incorporated in the ICDs and low emission by the phones, as well as the small size of the magnets in the smartphones tested. STUDY REGISTRATION: NCT02330900 (http://www.clinicaltrials.gov). Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author 2016. For permissions please email: journals.permissions@oup.com.

Entities:  

Keywords:  Electromagnetic interference; Implantable cardioverter defibrillator; Magnet effect; Mobile phone; Smartphone

Mesh:

Year:  2016        PMID: 26857191     DOI: 10.1093/europace/euv374

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  3 in total

Review 1.  [Electromagnetic interference : Pacemakers, cardiac resynchronization therapy devices, implantable cardioverter-defibrillator].

Authors:  Bernd Nowak; Oliver Przibille; Andreas Napp
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2022-07-04

Review 2.  Interference by Modern Smartphones and Accessories with Cardiac Pacemakers and Defibrillators.

Authors:  Fahd Nadeem; Cao Thach Tran; Estelle Torbey; Daniel Philbin; Carlos Morales; Michael Wu
Journal:  Curr Cardiol Rep       Date:  2022-01-27       Impact factor: 2.931

3.  Effect of ElectroMagnetic interference from SmartPHone on cardiac ImplaNtable electronic device (EMI-PHONE study).

Authors:  Sanatcha Apakuppakul; Nilubon Methachittiphan; Sirin Apiyasawat
Journal:  J Arrhythm       Date:  2022-07-12
  3 in total