| Literature DB >> 26852245 |
B Ten Brinke1, N M C Mathijssen1, I Blom2, R L M Deijkers3, E M Ooms4, G A Kraan1.
Abstract
The primary aim of this clinical and radiostereometric study was to study the migration pattern of the surface replacement trapeziometacarpal joint prosthesis (SRTMTMC, Avanta®, San Diego, CA). The secondary aims were to assess patient-related outcomes and prosthesis survival 5 years after surgery. Ten patients received the prosthesis. Radiostereometric radiographs were obtained 6 weeks, 6 months, 1 year and 5 years post-operatively and were analysed using model-based software. All patients completed DASH and Nelson Hospital scores at these follow-ups. Mean translations varied between 0.0 and 0.5 mm after 5 years. Rotation values could be calculated in six patients and mean rotations varied between -0.3 and 2.3°, although the precision of rotation values seems to be poor. The 5-year survival rate was 80%. Mean pre-operative DASH and Nelson Hospital scores were 53 (SD 14) and 51 (SD 13), respectively. Six months post-operatively, the DASH and Nelson Hospital scores had both significantly improved to 25 (SD 20) and 74 (SD 18) and remained high after 5 years. Implant stability was good 5 years post-operatively, and early migration did not predict implant failure in this study. LEVEL OF EVIDENCE: IV.Entities:
Keywords: Roentgen stereophotogrammetric analysis; migration; prosthesis survival; total joint arthroplasty; trapeziometacarpal joint
Mesh:
Year: 2016 PMID: 26852245 DOI: 10.1177/1753193416629070
Source DB: PubMed Journal: J Hand Surg Eur Vol ISSN: 0266-7681