Literature DB >> 26849029

Failure analysis of retrieved PE-UHMW acetabular liners.

Anna Laska1, Vasiliki-Maria Archodoulaki2, Bernadette Duscher3.   

Abstract

Ultra-high molecular weight polyethylene (PE-UHMW) acetabular liners have a limited lifespan in a patient's body. There are many factors affecting the performance of the implant and furthermore the properties of the polymeric material are changing after implantation. In this work material changes according to structure and morphology and their implication on mechanical properties are in focus. The physical and mechanical properties of ten crosslinked (xL) PE-UHMW and nine conventional (conv) gamma-sterilized PE-UHMW hip components, used as sliding surface in total hip joint replacement, with different in-vivo times are compared. The evaluation of the retrieved acetabular liners is performed in view of crosslinking and conventional gamma-sterilization but also in terms of the influence of gender concerning alteration in properties. The oxidative degradation in the PE-UHMW is investigated by means of Fourier Transformed Infrared Spectroscopy (FTIR). The characterization of the morphology is carried out via differential scanning calorimetry (DSC). A depth profile of the micro-hardness and elastic modulus is taken over the cross-section of the components in order to find the influence of chemical constitution and morphology on the micro-mechanical properties. It could be shown that crosslinking and oxidative degradation influence the degree of crystallinity of the polymer. Oxidation occurs for both types of the material due to in-vivo time. Higher degree of crystallinity can be correlated to higher hardness and indentation modulus. No unequivocal superiority of crosslinked over conventional liners can be observed. The influence of sex concerning alteration of the evaluated properties matters but need to be further investigated.
Copyright © 2016 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26849029     DOI: 10.1016/j.jmbbm.2016.01.007

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  1 in total

1.  Chemical stability of oil-infused polyethylene.

Authors:  Fedra P Zaribaf; Harinderjit S Gill; Elise C Pegg
Journal:  J Biomater Appl       Date:  2020-12-26       Impact factor: 2.646

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.