Literature DB >> 26845139

A Human Factors Engineering Study of the Medication Delivery Process during an Anesthetic: Self-filled Syringes versus Prefilled Syringes.

Yushi Yang1, Antonia Joy Rivera, Christopher R Fortier, James H Abernathy.   

Abstract

BACKGROUND: Prefilled syringes (PFS) have been recommended by the Anesthesia Patient Safety Foundation. However, aspects in PFS systems compared with self-filled syringes (SFS) systems have never been explored. The aim of this study is to compare system vulnerabilities (SVs) in the two systems and understand the impact of PFS on medication safety and efficiency in the context of anesthesiology medication delivery in operating rooms.
METHODS: This study is primarily qualitative research, with a quantitative portion. A work system analysis was conducted to analyze the complicated anesthesia work system using human factors principles and identify SVs. Anesthesia providers were shadowed: (1) during general surgery cases (n = 8) exclusively using SFS and (2) during general surgery cases (n = 9) using all commercially available PFS. A proactive risk assessment focus group was followed to understand the risk of each identified SV.
RESULTS: PFS are superior to SFS in terms of the simplified work processes and the reduced number and associated risk of SVs. Eight SVs were found in the PFS system versus 21 in the SFS system. An SV example with high risk in the SFS system was a medication might need to be "drawn-up during surgery while completing other requests simultaneously." This SV added cognitive complexity during anesthesiology medication delivery. However, it did not exist in the PFS system.
CONCLUSIONS: The inclusion of PFS into anesthesiology medication delivery has the potential to improve system safety and work efficiency. However, there were still opportunities for further improvement by addressing the remaining SVs and newly introduced complexity.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26845139     DOI: 10.1097/ALN.0000000000001040

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  6 in total

1.  ISMP Medication Error Report Analysis: Aggrastat-Argatroban Mix-ups Don't Expect Radiofrequency Identification Stock Systems To Be Perfect Paralyzed by Mistakes: Reassess the Safety of Neuromuscular Blockers in Your Facility.

Authors:  Michael R Cohen; Judy L Smetzer
Journal:  Hosp Pharm       Date:  2016-12

2.  Work systems analysis of sterile processing: assembly.

Authors:  Myrtede Alfred; Ken Catchpole; Emily Huffer; Larry Fredendall; Kevin M Taaffe
Journal:  BMJ Qual Saf       Date:  2020-10-19       Impact factor: 7.035

3.  Work systems analysis of sterile processing: decontamination.

Authors:  Myrtede Alfred; Ken Catchpole; Emily Huffer; Larry Fredendall; Kevin M Taaffe
Journal:  BMJ Qual Saf       Date:  2019-11-13       Impact factor: 7.035

4.  Interactional Resources for Quality Improvement: Learning From Participants Through a Qualitative Study.

Authors:  Joanna Veazey Brooks; Ksenia Gorbenko; Charles Bosk
Journal:  Qual Manag Health Care       Date:  2017 Apr/Jun       Impact factor: 0.926

5.  Intubation drug pack containing pre-filled syringes reduces the time to endotracheal intubation in a simulated paediatric emergency.

Authors:  Sara Lawson; Christopher Flannigan; Carolyn Green; Lynsey Freeburn; Anne Burns; Joseph McCann; Thomas Bourke
Journal:  BMJ Simul Technol Enhanc Learn       Date:  2020-03-02

6.  A national survey on attitudes and barriers on recycling and environmental sustainability efforts among Canadian anesthesiologists: an opportunity for knowledge translation.

Authors:  Maria-Alexandra Petre; Lisa Bahrey; Mark Levine; Adriaan van Rensburg; Mark Crawford; Clyde Matava
Journal:  Can J Anaesth       Date:  2018-12-13       Impact factor: 5.063

  6 in total