Extension of the standard addition method by blank addition.

Panagiotis Steliopoulos1.   

Abstract

Standard addition involves adding varying amounts of the analyte to sample portions of fixed mass or fixed volume and submitting those portions to the sample preparation procedure. After measuring the final extract solutions, the observed signals are linearly regressed on the spiked amounts. The original unknown amount is estimated by the opposite of the abscissa intercept of the fitted straight line [1]. A limitation of this method is that only data points with abscissa values equal to and greater than zero are available so that there is no information on whether linearity holds below the spiking level zero. An approach to overcome this limitation is introduced.•Standard addition is combined with blank addition.•Blank addition means that defined mixtures of blank matrix and sample material are subjected to sample preparation to give final extract solutions.•Equations are presented to estimate the original unknown amount and to calculate the 1-2α confidence interval about this estimate using the combined data set.

Entities:  

Keywords:  Blank addition; Chemical analysis; Quantification; Standard addition

Year:  2015        PMID: 26844210      PMCID: PMC4703584          DOI: 10.1016/j.mex.2015.09.001

Source DB:  PubMed          Journal:  MethodsX        ISSN: 2215-0161


Method details

Suppose that a liquid sample contains a particular compound of interest at unknown concentration. To determine this concentration, n aliquots of the sample are spiked with the analyte at concentration levels . Furthermore, sample material is diluted with blank matrix to give mixtures with sample volume fractions of (ratios of sample volume to volume of total mixture). From each solution of the two series an aliquot of the same volume is taken and submitted to chemical analysis. The observed measurement values y1, …, y are considered as realizations of random variables Y1, …, Y. Our statistical model iswhereandx* designates the unknown initial concentration of the analyte. The parameter β1 is the slope and the parameter β0 is the ordinate intercept of the straight line that relates the expectation of the signal to the spiked concentration. Note that β0 is equal to β1x*. Put another way, the functional relationship between the expectation of the signal and the totally present concentration is supposed to be given by a straight line through the origin. ɛ is the vector of the measurement errors. We assume that ɛ1, …, ɛ is a sequence of independent and identically normally distributed random variables with expectation value zero and variance σ2. Minimizing the sum of the squared deviations provides an estimator b for the parameter vector β:that isandwith i = 1, …, n and j = 1, …, m (b1 is an estimator for the slope β1 and b0 is an estimator for the ordinate intercept β0). An estimator for the originally present unknown concentration is Note that by replacing the random variables Y1, …, Y with the corresponding observed values y1, …, y, the above estimators become estimates. The abscissa values of the blank addition data points can be computed as Un unbiased estimator for the variance σ2 is Un unbiased estimator for the variance of the estimator b0 + b1x0 at a certain abscissa value x0 is The limits of a 1 − 2α confidence interval for β0 + β1x0 arewhere t1− denotes the 1 − α quantile of the t-distribution with n + m − 2 degrees of freedom. Setting Lupper/lower equal to zero and solving for x0 yields the limits of a 1 − 2α confidence interval for the abscissa intercept . The corresponding opposites are the limits of a 1 − 2α confidence interval for the unknown initial concentration :with The advantage of the extended approach described in this paper over the classical standard addition method [1] is that the former allows to see whether linearity holds below the spiking level zero. For illustration, consider the following example. Suppose that the expected signal is related to the present concentration by the curve shown in Fig. 1(a). Analysis of a sample which contains the analyte at a concentration of 2 μg/L using the standard addition method gives the line depicted in Fig. 1(b) and a result of 7.0 μg/L. Although the standard addition plot looks quite good, the result considerably exceeds the actual concentration. Combining standard addition with blank addition (Fig. 1(c)) leads to a result of 8.8 μg/L, which is even higher, but now it is obvious that the outcome is erroneous because the blank addition data points clearly deviate from the fitted straight line and follow a curve. A basic assumption of the combined approach is that the sample matrix and the added blank matrix cause the same matrix effect; or, mathematically speaking, the relationship between the expected signal and the present concentration for both matrices is given by the same line through the origin. Fig. 2 shows two situations in which the model Eq. (1) is fitted to a standard and blank addition data set that does not fulfill this requirement. As it can be seen from this figure, when the slope that relates the expected signal to the present concentration is greater in the case of the sample matrix than in the case of the matrix used for blank addition, then the combined approach will tend to overestimate the unknown initial concentration. Conversely, when the slope that relates the expected signal to the present concentration is smaller in the case of the sample matrix than in the case of the matrix used for blank addition, then the combined approach will tend to underestimate the unknown initial concentration. The application of the combined approach is not only restricted to liquid matrices. Some examples from the laboratory of the author where the combined approach proved to be suitable are: quantification of steroids in urine by LC–MS, quantification of antibiotic residues in animal tissues (muscle and kidney) by LC–MS and quantification of artificial sweeterners in mineral water by LC–MS. Table 1 and Fig. 3 summarize the results of the analysis of a male bovine urine sample for its content of the steroid nandrolone by an LC–MS assay. A series of six aliquots of the urine sample were spiked with the target analyte at levels of 0, 0.5, 1.0, 1.5, 2.0 and 2.5 μg/L. Additionally, dilutions of the urine sample with nandrolone-free urine were prepared at sample volume fractions of 0.1, 0.3, 0.5, 0.7 and 0.9. Table 1 lists the observed signals, the estimated parameters, the estimate for the originally present unknown concentration, the 1 − 2α confidence interval for the originally present unknown concentration and the calculated abscissa values of the blank addition data points. For comparison, both the combined approach and the standard addition method were applied to quantify nandrolone. The two methods provided similar estimates. The smaller confidence interval obtained with the extended method is due to the higher number of data points used. Fig. 3 displays the data points, the straight line fitted to the combined data set and the 0.95 confidence bands.
Fig. 1

An illustration of what might happen if nonlinearity occurs below the spiking level zero.

Assumed nonlinear relationship between the expected signal and the present concentration.

Standard addition.

Standard addition combined with blank addition.

Fig. 2

An illustration of what might happen if standard addition is combined with blank addition, although the sample matrix and the added blank matrix exhibit different matrix effects.

The slope that relates the expected signal to the present concentration is greater in the case of the sample matrix than in the case of the matrix used for blank addition.

The slope that relates the expected signal to the present concentration is smaller in the case of the sample matrix than in the case of the matrix used for blank addition.

Table 1

Quantification of nandrolone in a bovine urine sample.

Data
Quantification by standard addition combined with blank addition
Quantification by standard addition
xiyikiyn+jb0b1xˆn+jxˆ*0.95 confidence intervalb0b1xˆ*0.95 confidence interval
0.053947080.146413949500602696536−1.651.84[1.61; 2.10]414024725927971.98[1.56; 2.54]
0.560617050.31497300−1.29
1.078727400.52492913−0.92
1.588223560.73069726−0.55
2.0105483850.94348496−0.18
2.511587569

n = 6, m = 5, combined approach: t1−0.25 = 2.26 (9 degrees of freedom), standard addition: t1−0.25 = 2.78 (4 degrees of freedom).

Fig. 3

Quantification of nandrolone in a bovine urine sample by standard addition combined with blank addition (plot of the standard and blank addition line, the 0.95 confidence bands and the data points).

  2 in total

1.  Ionic Exchange Resins and Hydrogels for Capturing Metal Ions in Selected Sweet Dessert Wines.

Authors:  Gabriella Tamasi; Alessio Pardini; Claudia Bonechi; Alessandro Donati; Mario Casolaro; Gemma Leone; Marco Consumi; Renzo Cini; Agnese Magnani; Claudio Rossi
Journal:  Molecules       Date:  2018-11-14       Impact factor: 4.411

2.  Supercritical Fluid Chromatography with Photodiode Array Detection in the Determination of Fat-Soluble Vitamins in Hemp Seed Oil and Waste Fish Oil.

Authors:  Katarzyna Tyśkiewicz; Roman Gieysztor; Izabela Maziarczyk; Paweł Hodurek; Edward Rój; Krystyna Skalicka-Woźniak
Journal:  Molecules       Date:  2018-05-10       Impact factor: 4.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.