Literature DB >> 26840919

Validation of an automated enzyme immunoassay for the measurement of serum total thyroxine in cats.

Tim L Williams1, Joy Archer1.   

Abstract

BACKGROUND: Hyperthyroidism is common in older cats, which necessitates frequent screening of serum total thyroxine (TT4) concentrations. Fast, cheap, and reliable methods to measure TT4 in cats are needed.
OBJECTIVES: The purpose of the study was the validation of a human TT4 enzyme immunoassay (EIA) for use with feline serum, and derivation of a TT4 reference interval (RI) for cats aged 9 years and older.
METHODS: Assay precision, reproducibility, and linearity were evaluated. Interference by hemolysis was also assessed. Method comparison studies between the human EIA and a previously validated radioimmunoassay (RIA) and chemiluminescent-enzyme immunoassay (CEIA) were performed. Healthy cats (> 9 years) were recruited from 3 UK first opinion practices.
RESULTS: The human TT4 EIA demonstrated good precision and reproducibility, and adequate linearity. Hemolysis did not significantly alter measured TT4 concentrations until HGB > 8 g/L. Method comparison revealed proportional and constant errors between EIA and RIA/CEIA. The TT4 RI for cats (> 9 years) was calculated as 7.1-45.1 nmol/L (n = 49).
CONCLUSIONS: The human TT4 EIA was successfully validated for use with feline serum and offers a rapid, cheap, and reliable method for determination of serum TT4 concentrations in cats.
© 2016 American Society for Veterinary Clinical Pathology.

Entities:  

Keywords:  Analyzer; feline; human; hyperthyroidism; reference interval

Mesh:

Substances:

Year:  2016        PMID: 26840919     DOI: 10.1111/vcp.12324

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  3 in total

1.  Serum Cystatin C Concentrations in Cats with Hyperthyroidism and Chronic Kidney Disease.

Authors:  T L Williams; H Dillon; J Elliott; H M Syme; J Archer
Journal:  J Vet Intern Med       Date:  2016-05-07       Impact factor: 3.333

2.  Comparison of 2 assays for measuring serum total thyroxine concentration in dogs and cats.

Authors:  Ewan D S Wolff; Graham Bilbrough; George Moore; Lynn Guptill; J Catharine Scott-Moncrieff
Journal:  J Vet Intern Med       Date:  2020-02-03       Impact factor: 3.333

3.  Correlation of thyroid hormone measurements with thyroid stimulating hormone stimulation test results in radioiodine-treated cats.

Authors:  Jennifer Wakeling; Teresa Hall; Timothy L Williams
Journal:  J Vet Intern Med       Date:  2020-10-12       Impact factor: 3.333

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.