Literature DB >> 26826941

Retrospective Review of 516 Implantations of Option Inferior Vena Cava Filters at a Single Health Care System.

Brian Tsui1, Tianzhi An2, Eunice Moon2, Russell King2, Weiping Wang3.   

Abstract

PURPOSE: To retrospectively evaluate the safety, efficacy, and retrievability of Option inferior vena cava (IVC) filters.
MATERIALS AND METHODS: All patients (N = 516; 247 women; mean age, 67.1 y ± 15.1; range, 19.5-101.6 y) who received an Option filter between August 2009 and March 2015 at a single health care system were analyzed.
RESULTS: The study duration was 68 months, with median clinical follow-up of 7.1 months (range, 1 d to 61.8 mo). During follow-up, 73 of 83 patients (88.0%) underwent successful filter retrieval, 153 died (including three after successful retrieval), and 293 remained alive with filters in situ. Seventeen cases of breakthrough pulmonary embolism (PE) occurred (3.4%). Among 323 patients with direct filter imaging, there were two cases of tilt > 15°, one case of filter deformity, 16 cases of intracaval migration > 2 cm, and no cases of filter fracture. There were six cases of caval occlusion, nine cases of thrombus trapped inside the filter, and 57 cases of limb penetration on computed tomography scans or radiographs of the IVC. Retrieval failures were attributed to filter tilt or tip embedment in the caval wall (n = 4), complete IVC thrombosis (n = 3), thrombus inside the filter (n = 2), or inability to disengage filter legs (n = 1). Recurrent deep vein thrombosis occurred in 34 patients, including 32 with filters in situ and two whose filters had been removed.
CONCLUSIONS: Most Option filters were left in situ for permanent indications. Rates of successful retrieval, device-related complications, and breakthrough PE were similar to those associated with other retrievable filters.
Copyright © 2016 SIR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 26826941     DOI: 10.1016/j.jvir.2015.11.055

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  3 in total

1.  An analysis of factors associated with increased fluoroscopy time or the need for complex techniques at IVC filter retrieval.

Authors:  Mark Kleedehn; Kelli Moore; Katherine Longo; Kaitlin Woo; Paul Laeseke
Journal:  Eur Radiol       Date:  2018-10-09       Impact factor: 5.315

Review 2.  Over-the-Wire Inferior Vena Cava Filter Placement: How We Do It.

Authors:  Xin Li; Jennifer Montgomery; Levester Kirksey; Sameer Gadani; Giuseppe D'Amico; Sasan Partovi
Journal:  Semin Intervent Radiol       Date:  2021-06-03       Impact factor: 1.780

3.  The role of an IVC filter retrieval clinic-A single center retrospective analysis.

Authors:  Philip A Schuchardt; Junaid T Yasin; Ryan M Davis; Sanjit O Tewari; Ambarish P Bhat
Journal:  Indian J Radiol Imaging       Date:  2019-12-31
  3 in total