Mamdouh R Rezk1, Emad B Basalious2, Mohammed E Amin3.
Abstract
A novel and sensitive LC-MS/MS method was developed and validated for determination of sofosbuvir (SF) using eplerenone as an internal standard. The Xevo TQD LC-MS/MS was operated under the multiple-reaction monitoring mode using electrospray ionization. Extraction with tert-butyl methyl ether was used in sample preparation. The prepared samples were chromatographed on Acquity UPLC BEH C18 (50 × 2.1 mm, 1.7 μm) column by pumping 0.1% formic acid and acetonitrile in an isocratic mode at a flow rate of 0.35 mL/min. Method validation was performed as per the US Food and Drug Administration guidelines and the standard curves were found to be linear in the range of 0.25-3500 ng/mL for SF. The intra- and inter-day precision and accuracy results were within the acceptable limits. A very short run time of 1 min made it possible to analyze more than 500 human plasma samples per day. A very low quantification limit of SF allowed the applicability of the developed method for determination of SF in a bioequivalence study in human volunteers.Entities:
Keywords: UPLC-MS/MS; bioequivalence; plasma; sofosbuvir; validation
Mesh:
Substances:
Year: 2016 PMID: 26821881 DOI: 10.1002/bmc.3690
Source DB: PubMed Journal: Biomed Chromatogr ISSN: 0269-3879 Impact factor: 1.902