| Literature DB >> 26816509 |
Arpita Bindal1, Umesh Mahantshetty1, Chandrashekhar M Tambe2, Yogesh Ghadi2, Vedang Murthy1, Shyam Kishore Shrivastava1.
Abstract
PURPOSE: Penile cancer, although rare, is one of the common genitourinary cancers in India affecting mostly aged uncircumcised males. For patients presenting with small superficial lesions < 3 cm restricted to glans, surgery, radical external radiation or brachytherapy may be offered, the latter being preferred as it allows organ and function preservation. In patients receiving brachytherapy, testicular morbidity is not commonly addressed. With an aim to minimize and document the doses to testis after adequate shielding during radical interstitial brachytherapy for penile cancers, we undertook this study in 2 patients undergoing brachytherapy and forms the basis of this report.Entities:
Keywords: HDR brachytherapy; penile brachytherapy; radiation; shielding; testicular cancer
Year: 2015 PMID: 26816509 PMCID: PMC4716126 DOI: 10.5114/jcb.2015.56410
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Schematic representation of thermoluminescent dosimeters placement
Fig. 2Patient A without (A) and with (B) lead shield in place
Dosimetric uncertainties for 192Ir brachytherapy source
| Source of uncertainty | Value | Reference |
|---|---|---|
| Air Kerma Strength measurement | 1.5% | [ |
| Source to TLD positioning | < 3% | [ |
| TLD calibration using 60Co | 3% |
[ |
| Temporal fading | Negligible | All readouts were done at 24 h from irradiation |
| TLD readout uncertainties | 2% | Including PMT linearity correction, reader stability, and TLD reproducibility (institutional data) |
| Type A uncertainties | < 3% | |
| Total dosimetric uncertainty | 6% | Coverage factor [k] = 1 |
PMT – photomultiplier tube, TLD – thermoluminescent dosimeter
Actual thermoluminescent dosimeters readings and calculated parameters obtained for the two patients
| Patient A (D = 4400 cGy/11#) | Patient B (D = 4800 cGy/12#) | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| R1 | R2 | R3 | Mean | SD | Total (in cGy) | R1 | R2 | R3 | Mean | SD | Total (in cGy) | |
|
| 64.2 | 72.34 | 60.4 | 65.65 | 6.10 | 722.15 ± 67.10 | 73.3 | 62.14 | 66.50 | 67.31 | 5.62 | 807.72 ± 67.44 |
|
| 3.44 | 3.69 | 3.10 | 3.41 | 0.29 | 37.51 ± 3.19 | 2.84 | 1.61 | 2.25 | 2.23 | 0.61 | 26.76 ± 7.32 |
|
| 2.19 | 1.96 | 2.90 | 2.35 | 0.49 | 25.85 ± 5.39 | 1.45 | 1.14 | 1.67 | 1.42 | 0.26 | 17.04 ± 3.12 |
|
| 2.82 | 2.83 | 3.00 | 2.88 | 0.10 | 31.68 ± 1.10 | 2.15 | 1.38 | 1.96 | 1.83 | 0.40 | 21.96 ± 4.80 |
|
| 6.64 | 3.74 | 4.90 | 5.09 | 1.45 | 55.99 ± 15.95 | 2.57 | 2.29 | 2.17 | 2.34 | 0.20 | 28.08 ± 2.40 |
|
| 2.45 | 2.41 | 3.20 | 2.69 | 0.44 | 29.59 ± 4.84 | 1.42 | 1.65 | 1.55 | 1.54 | 0.11 | 18.48 ± 1.32 |
|
| 4.55 | 3.08 | 4.05 | 3.89 | 0.74 | 42.79 ± 8.14 | 2.00 | 1.97 | 1.86 | 1.94 | 0.07 | 23.28 ± 0.84 |
Readings R1, R2, and R3 obtained on TLD 1 (Penis Post; ventral surface of penile shaft), TLD 2 (Lt Ant; anterior surface of left testis), TLD 3 (Lt Post; posterior surface of left testis), TLD 4 (Rt Ant; anterior surface of right testes), and TLD 5 (Rt Post; posterior surface of right testes). Lt Mid Calc and Rt Mid Calc denote calculated mean of the doses to the anterior and posterior surfaces of the left and right testes, respectively, as measured in 3 fractions.