Increased Computed Tomography Dose Due to Miscentering With Use of Automated Tube Voltage Selection: Phantom and Patient Study.

Peter D Filev1, Pardeep K Mittal1, Xiangyang Tang1, Phuong-Anh Duong1, Xiaojing Wang2, William C Small1, Kimberly Applegate1, Courtney C Moreno3.   

Abstract

The purpose of the article is to determine if miscentering affected dose with use of automated tube voltage selection software. An anthropomorphic phantom was imaged at different table heights (centered in the computed tomography [CT] gantry, and -6, -3, +3, and +5.7cm relative to the centered position). Topogram magnification, tube voltage selection, and dose were assessed. Effect of table height on dose also was assessed retrospectively in human subjects (n = 50). When the CT table was positioned closer to the x-ray source, subjects appeared up to 33% magnified in topogram images. When subjects appeared magnified in topogram images, automated software selected higher tube potentials and tube currents that were based on the magnified size of the subject rather than the subject׳s true size. Table height strongly correlated with CT dose index (r = 0.98, P < 0.05) and dose length product (r = 0.98, P < 0.05) in the phantom study. Transverse dimension in the topogram highly correlated with dose in human subjects (r = 0.75-0.87, P <0.05). Miscentering results in increased dose due to topogram magnification with automated voltage selection software.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2015        PMID: 26810714     DOI: 10.1067/j.cpradiol.2015.11.003

Source DB:  PubMed          Journal:  Curr Probl Diagn Radiol        ISSN: 0363-0188


  2 in total

1.  A CONSORT-compliant prospective randomized controlled trial: radiation dose reducing in computed tomography using an additional lateral scout view combined with automatic tube current modulation: Phantom and patient study.

Authors:  Wanlin Peng; Zhenlin Li; Chunchao Xia; Yingkun Guo; Jinge Zhang; Kai Zhang; Lei Li; Fei Zhao
Journal:  Medicine (Baltimore)       Date:  2017-07       Impact factor: 1.889

2.  The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.

Authors:  Touko Kaasalainen; Teemu Mäkelä; Mika Kortesniemi
Journal:  Eur J Radiol Open       Date:  2018-12-17
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.