| Literature DB >> 26809084 |
Ishanth Devinda Jayewardene1, Ashleigh Xie, Arjun Iyer, Roger Pye, Kumud Dhital.
Abstract
A limitation of veno-venous extracorporeal membrane oxygenation (VV-ECMO) is the recirculating of newly oxygenated blood back to the VV-ECMO circuit. We compared cannulation strategies for combating recirculation utilizing a mock circulation loop (MCL) with ultrasonic flow probes and pressure transducers ensuring that the MCL was run at physiological hemodynamic parameters. Mean recirculation percentages were lower (ANOVA, F = 14.25; p = 0.0001) with the Wang-Zwische dual lumen cannula (4.00 ± 1.77, n = 7) than both the femoro-jugular (15.23 ± 7.00, n = 8) and the femoro-femoral cannulation configurations (13.49 ± 1.44, n = 8).Entities:
Mesh:
Year: 2016 PMID: 26809084 DOI: 10.1097/MAT.0000000000000350
Source DB: PubMed Journal: ASAIO J ISSN: 1058-2916 Impact factor: 2.872