Literature DB >> 26809084

Development of a Mock Extracorporeal Membrane Oxygenation Circuit to Assess Recirculation.

Ishanth Devinda Jayewardene1, Ashleigh Xie, Arjun Iyer, Roger Pye, Kumud Dhital.   

Abstract

A limitation of veno-venous extracorporeal membrane oxygenation (VV-ECMO) is the recirculating of newly oxygenated blood back to the VV-ECMO circuit. We compared cannulation strategies for combating recirculation utilizing a mock circulation loop (MCL) with ultrasonic flow probes and pressure transducers ensuring that the MCL was run at physiological hemodynamic parameters. Mean recirculation percentages were lower (ANOVA, F = 14.25; p = 0.0001) with the Wang-Zwische dual lumen cannula (4.00 ± 1.77, n = 7) than both the femoro-jugular (15.23 ± 7.00, n = 8) and the femoro-femoral cannulation configurations (13.49 ± 1.44, n = 8).

Entities:  

Mesh:

Year:  2016        PMID: 26809084     DOI: 10.1097/MAT.0000000000000350

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872