Literature DB >> 26800173

Influence of Platform-Switched, Laser-Microtextured Implant on Marginal Bone Level: A 24-Month Case Series Study.

Vincenzo Iorio-Siciliano, Gaetano Marenzi, Andrea Blasi, Jolanda Mignogna, Carlo Cafiero, Hom-Lay Wang, Gilberto Sammartino.   

Abstract

PURPOSE: The aim of this case series study was to evaluate clinical and radiographic changes of soft and hard tissues around tapered, platform-switched, laser-microtextured implants 24 months after crown placement.
MATERIALS AND METHODS: Twenty tapered, platform-switched, laser-microtextured collar implants were placed in 20 patients. Full-mouth plaque score, full-mouth bleeding score, probing depth, and mucosal recession were recorded at the time of crown cementation and after 24 months follow-up. The marginal bone-level changes at the mesial and distal aspects of the implants were calculated by subtracting from baseline and 24-month implant marginal bone level.
RESULTS: In terms of the full-mouth plaque score and full-mouth bleeding score, tapered, platform-switched, laser-microtextured implants showed statistically significant improvements at 6 months when compared to baseline (P < .001). Statistically significantly deeper probing depths (P < .001) were found when comparing baseline and at 24 months at mesial, lingual, and distal sites. However, no statistically significant difference was found at the buccal aspects (P = .064). Radiographic marginal bone loss at 2-year follow-up for tapered, platform-switched, laser-microtextured implants was 0.72 ± 0.16 mm and 0.67 ± 0.15 mm at the mesial and distal sites, respectively.
CONCLUSION: Within the limits of this study, tapered, platform-switched, laser-microtextured implants maintained marginal bone level (less than 1 mm radiographic bone loss) as well as limited mucosa recession over a 2-year period.

Entities:  

Mesh:

Year:  2016        PMID: 26800173     DOI: 10.11607/jomi.4130

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  3 in total

1.  Does the Laser-Microtextured Short Implant Collar Design Reduce Marginal Bone Loss in Comparison with a Machined Collar?

Authors:  B Alper Gultekin; Ali Sirali; Pinar Gultekin; Serdar Yalcin; Eitan Mijiritsky
Journal:  Biomed Res Int       Date:  2016-08-31       Impact factor: 3.411

2.  Immediate implant placement in anterior teeth with grafting material of autogenous tooth bone vs xenogenic bone.

Authors:  Dong Wu; Lin Zhou; Jichao Lin; Jiang Chen; Wenxiu Huang; Yonghui Chen
Journal:  BMC Oral Health       Date:  2019-12-02       Impact factor: 2.757

3.  Interaction between Different Implant Surfaces and Liquid Fibrinogen: A Pilot In Vitro Experiment.

Authors:  Catherine X Andrade; Marc Quirynen; David R Rosenberg; Nelson R Pinto
Journal:  Biomed Res Int       Date:  2021-07-05       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.