Pulse Oximetry Overestimates Oxyhemoglobin in Neonates with Critical Congenital Heart Disease.

Daniel Murphy1, Youngju Pak, John Patrick Cleary.   

Abstract

BACKGROUND: Pulse oximetry is a key part of the clinical evaluation and management of neonates with congenital heart defects. In 2011, the US Department of Health and Human Services recommended use of routine pulse oximetry to screen for critical congenital heart disease (CCHD). Current studies suggest pulse oximetry overestimates arterial oxygen saturation in moderately hypoxemic pediatric patients. Based on variable hypoxemia in neonates with CCHD, concern exists that present pulse oximeter technology may overestimate measured oxyhemoglobin.
OBJECTIVES: To compare pulse oximetry and oxyhemoglobin values in NICU patients with known CCHD to evaluate the ability of pulse oximetry to reliably predict oxyhemoglobin accounting for potential confounding variables such as heart lesion, saturation range, total hemoglobin concentration, peripheral perfusion, and timing of measurements.
METHODS: This is a single-center retrospective study. Inclusion criteria were AHA-defined CCHD and umbilical artery blood gas-derived oxyhemoglobin with concurrent pulse oximetry recording during hours of life 0-72. Bland-Altman analysis and the concordance correlation coefficient were used to measure the internal consistency (agreement) between the two measurements.
RESULTS: 89 patients were evaluated with 599 paired arterial oxyhemoglobin and pulse oximetry recordings. 47% of all pulse oximetry values were ≥95% - the cutoff for CCHD screening. Pulse oximetry overestimated arterial oxyhemoglobin by a mean of 5.4% over all levels of oxygen saturation. Pulse oximetry overestimation was >3 in 65.4% of measurements, >6 in 41.2% of measurements, and >10 in 15.3% of measurements. Hour of life, total hemoglobin, and peripheral perfusion did not significantly affect the degree of overestimation.
CONCLUSIONS: Our results reinforce the concern that present pulse oximeters overestimate oxyhemoglobin values, contributing to some false-negative CCHD screens and potentially leading to unnecessary escalations in care of those patients with diagnosed CCHD. Improvements in pulse oximetry accuracy and precision in the neonate would benefit both screening and clinical care in the NICU.
© 2016 S. Karger AG, Basel.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26800085     DOI: 10.1159/000442379

Source DB:  PubMed          Journal:  Neonatology        ISSN: 1661-7800            Impact factor:   4.035


  2 in total

1.  Development and Validation of Extracorporeal Membrane Oxygenation Mortality-Risk Models for Congenital Diaphragmatic Hernia.

Authors:  Yigit S Guner; Danh V Nguyen; Lishi Zhang; Yanjun Chen; Matthew T Harting; Peter Rycus; Ryan Barbaro; Matteo Di Nardo; Thomas V Brogan; John P Cleary; Peter T Yu
Journal:  ASAIO J       Date:  2018 Nov/Dec       Impact factor: 2.872

2.  Trends in Mortality and Risk Characteristics of Congenital Diaphragmatic Hernia Treated With Extracorporeal Membrane Oxygenation.

Authors:  Yigit S Guner; Patrick T Delaplain; Lishi Zhang; Matteo Di Nardo; Thomas V Brogan; Yanjun Chen; John P Cleary; Peter T Yu; Matthew T Harting; Henri R Ford; Danh V Nguyen
Journal:  ASAIO J       Date:  2019-07       Impact factor: 2.872

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.