Literature DB >> 26790794

Complications resulting from tracker pin-sites in computer navigated knee replacement surgery.

Anna Thomas, Gopalakrishna Pemmaraju, Gurpreet Nagra, James Bassett, Shreeram Deshpande.   

Abstract

Intra-osseous pin sites used in Computer Navigated knee Arthroplasty are known to cause complications including infection, neurovascular injury, thermal necrosis and peri-prosthetic fracture. Many studies have looked at ways of reducing these risks to improve results in computer-navigated surgery. We present our complication rates and techniques used to reduce complications in a retrospective study of 321 patients performed by a single surgeon in between 2009 and 2013. One patient was identified with a superficial pin site infection. No major complications were recorded during a mean follow up period of 31 months. Attention to detail in every aspect of pin insertion reduces the rates of both minor and major complications. In particular rates of peri-prosthetic fracture were shown to be lower when using a uni-cortical drilling technique compared with the use of bicortical drilling in other studies.

Entities:  

Mesh:

Year:  2015        PMID: 26790794

Source DB:  PubMed          Journal:  Acta Orthop Belg        ISSN: 0001-6462            Impact factor:   0.500


  1 in total

1.  Periprosthetic Fracture through a Unicortical Tracking Pin Site after Computer Navigated Total Knee Replacement.

Authors:  Matt Blue; Christian Douthit; Joel Dennison; Cyrus Caroom; Mark Jenkins
Journal:  Case Rep Orthop       Date:  2018-09-16