Literature DB >> 26789097

Catastrophic strut collapse with the Taylor Spatial Frame: preventing a disaster.

F Birkholtz1, N Ferreira2.   

Abstract

Circular external fixators, and in particular hexapod circular external fixators, are increasingly being used in orthopaedic traumatology and limb reconstruction. Their ability to provide three dimensional stability throughout the treatment period make these external fixators indispensable in the management of complex reconstructive work. This study has identified a unique mode of failure with the use of Taylor Spatial Frame Fast-Fx struts and termed this complication catastrophic strut collapse. In this article a lab model is used to demonstrate the consequences of catastrophic strut collapse and warn users of the importance of using the ID bands and locking nuts to prevent inadvertent fragment displacement.

Keywords:  Fast-Fx; Taylor Spatial Frame; collapse; complication; external fixator; hexapod; strut

Mesh:

Year:  2016        PMID: 26789097     DOI: 10.3109/03091902.2015.1126364

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  1 in total

1.  An Alternative Technique to Strut Change in Hexapod Circular External Fixator.

Authors:  M Rao; K Jamil; R Tasarib
Journal:  Malays Orthop J       Date:  2021-11
  1 in total