Radiation dose in chest CT: Assessment of size-specific dose estimates based on water-equivalent correction.

M Gabusi1, L Riccardi2, C Aliberti3, S Vio4, M Paiusco1.   

Abstract

PURPOSE: In the setting of thorax Computed Tomography (CT), the main purpose of this work is to quantify differences in Size-Specific Dose Estimate (SSDE), calculated on the basis of both effective diameter and water-equivalent diameter, accounting for patient size and tissue attenuation, respectively.
MATERIALS AND METHODS: An in-house software was developed to measure water-equivalent diameter and effective diameter on each CT slice, for 133 CT examinations. SSDE was calculated according to the Report of American Association of Physicists in Medicine Task Group (AAPM TG) 204 and 220. The ratio between effective and water-equivalent diameters was studied as function of cross-sectional air-to-body area ratio, for the slice located in the middle of the scan range.
RESULTS: When AAPM TG 220 prescriptions were applied, SSDE was mostly found larger than that obtained with AAPM TG 204 recommendations. On average, a difference of about 12% was observed, in spite of a considerable variability (from -18% to 53%). The ratio between effective and water-equivalent diameters ranged from 0.97 to 1.31, with a mean value of 1.15. Moreover, it was found considerably correlated with cross-sectional air-to-body ratio (Pearson's coefficient was 0.78 for women, 0.90 for men).
CONCLUSION: The discrepancy obtained by calculating SSDE on the basis of AAPM TG 204, instead of AAPM TG 220 Report, may vary substantially. Fluctuations were justified by the variability observed in the relative amount of low attenuating tissues in body, which was proved to be considerably correlated to the ratio between effective and water-equivalent diameters.
Copyright © 2016. Published by Elsevier Ltd.

Entities:  

Keywords:  Computed tomography; Dose; SSDE; Water-equivalent diameter

Mesh:

Substances:

Year:  2016        PMID: 26781588     DOI: 10.1016/j.ejmp.2015.12.008

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

1.  Asian consortium on radiation dose of pediatric cardiac CT (ASCI-REDCARD).

Authors:  Peter K T Hui; Hyun Woo Goo; Jing Du; Janice J K Ip; Suzu Kanzaki; Young Jin Kim; Supika Kritsaneepaiboon; Oktavia Lilyasari; Suvipaporn Siripornpitak
Journal:  Pediatr Radiol       Date:  2017-04-24

2.  Breast dose reduction for chest CT by modifying the scanning parameters based on the pre-scan size-specific dose estimate (SSDE).

Authors:  Masafumi Kidoh; Daisuke Utsunomiya; Seitaro Oda; Takeshi Nakaura; Yoshinori Funama; Hideaki Yuki; Kenichiro Hirata; Tomohiro Namimoto; Daisuke Sakabe; Masahiro Hatemura; Yasuyuki Yamashita
Journal:  Eur Radiol       Date:  2016-10-07       Impact factor: 5.315

3.  Size-Specific Dose Estimates Based on Water-Equivalent Diameter and Effective Diameter in Computed Tomography Coronary Angiography.

Authors:  Jian Xu; Xiangquan Wang; Huawei Xiao; Jianguo Xu
Journal:  Med Sci Monit       Date:  2019-12-06

4.  Comparison of Application Value of Different Radiation Dose Evaluation Methods in Evaluating Radiation Dose of Adult Thoracic and Abdominal CT Scan.

Authors:  Jimin He; Guanwei Dong; Yi Deng; Jun He; ZhiGang Xiu; Fanzi Feng
Journal:  Front Surg       Date:  2022-03-25
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.