| Literature DB >> 26780959 |
Diana S Pope1, Erik T Miller-Klein.
Abstract
Hospitals have complex soundscapes that create challenges to patient care. Extraneous noise and high reverberation rates impair speech intelligibility, which leads to raised voices. In an unintended spiral, the increasing noise may result in diminished speech privacy, as people speak loudly to be heard over the din. The products available to improve hospital soundscapes include construction materials that absorb sound (acoustic ceiling tiles, carpet, wall insulation) and reduce reverberation rates. Enhanced privacy curtains are now available and offer potential for a relatively simple way to improve speech privacy and speech intelligibility by absorbing sound at the hospital patient's bedside. Acoustic assessments were performed over 2 days on two nursing units with a similar design in the same hospital. One unit was built with the 1970s' standard hospital construction and the other was newly refurbished (2013) with sound-absorbing features. In addition, we determined the effect of an enhanced privacy curtain versus standard privacy curtains using acoustic measures of speech privacy and speech intelligibility indexes. Privacy curtains provided auditory protection for the patients. In general, that protection was increased by the use of enhanced privacy curtains. On an average, the enhanced curtain improved sound absorption from 20% to 30%; however, there was considerable variability, depending on the configuration of the rooms tested. Enhanced privacy curtains provide measureable improvement to the acoustics of patient rooms but cannot overcome larger acoustic design issues. To shorten reverberation time, additional absorption, and compact and more fragmented nursing unit floor plate shapes should be considered.Entities:
Mesh:
Year: 2016 PMID: 26780959 PMCID: PMC4918673 DOI: 10.4103/1463-1741.174377
Source DB: PubMed Journal: Noise Health ISSN: 1463-1741 Impact factor: 0.867
Categories of speech privacy and speech intelligibility determined by Speech Intelligibility Index (SII) and Articulation Index (AI)
| Speech privacy “Confidential” | Speech privacy “Normal” to “marginal” | Speech privacy “No privacy” |
|---|---|---|
| SII <0.10 | 0.10 <SII <0.45 | 0.45 <SII |
| AI <0.05 | 0.05 <AI <0.35 | 0.35 <AI |
| Speech intelligibility “Poor” | Speech intelligibility “Fair/Normal” | Speech intelligibility “Excellent” |
Average* sound absorption coefficient (ᾱ) for patient rooms in standard and refurbished nursing units with standard and Hush Curtains® in place
| Rooms | Standard curtain | Hush Curtain® | |
|---|---|---|---|
| Standard unit | |||
| Room A — four-bed | 0.24 (eight curtains) | 0.30 (12 curtains) | |
| Room B — one-bed | 0.23 (one curtain) | 0.28 (two curtains) | |
| Room C — one-bed | 0.24 (one curtain) | Not tested | |
| Room D — two-bed | 0.27 (four curtains) | 0.34 (five curtains) | |
| Room E — four-bed | 0.31 (eight curtains) | Not tested | |
| Refurbished unit | |||
| Room F — two-bed | 0.24 (four curtains) | Not tested | |
| Room G — two-bed | 0.17 ( four curtains) | 0.28 (eight curtains) | |
| Room H — one-bed | 0.23 (one curtain) | 0.28 (two curtains) | |
*FGI Design value of 0.15 is considered to be an average, acceptable sound absorption for patient rooms[, A larger number indicates that the room more efficiently absorbs reflected sound and will generally be perceived as being quieter
Speech Intelligibility Index (SII) and Articulation Index (AI) measures made in five patient rooms on the standard unit in three acoustic conditions: No curtain, standard curtain, and Hush Curtain®
| Standard unit Four-bed-occupied Room “A” | No curtain | Standard curtain | Hush Curtain® | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| SII | AI | SII | AI | SII | AI | |||||
| Patient in room | 0.640 | 0.564 | 0.607 | 0.521 | 0.524 | 0.426 | ||||
| Patient in room | 0.472 | 0.357 | 0.415 | 0.310 | 0.347 | 0.211 | ||||
| Patient in room | 0.520 | 0.413 | 0.442 | 0.332 | 0.305 | 0.196 | ||||
| Door | 0.288 | 0.182 | 0.262 | 0.157 | 0.158 | 0.082 | ||||
| Patient to nurse station | 0.025 | 0.006 | 0.000 | 0.000 | 0.000 | 0.000 | ||||
| Nurse station to patient | 0.089 | 0.043 | 0.025 | 0.006 | 0.032 | 0.006 | ||||
| One-bed “mirror” rooms — unoccupied | ||||||||||
| Room “B” | ||||||||||
| Doorway | 0.669 | 0.587 | 0.663 | 0.579 | 0.583 | 0.481 | ||||
| Hallway | 0.489 | 0.376 | 0.471 | 0.359 | 0.372 | 0.248 | ||||
| Patient to nurse station | 0.158 | 0.044 | 0.149 | 0.046 | 0.098 | 0.022 | ||||
| Nurse station to patient | 0.145 | 0.039 | 0.137 | 0.042 | 0.100 | 0.029 | ||||
| Room “C” | ||||||||||
| Doorway | 0.673 | 0.595 | 0.664 | 0.583 | ||||||
| Hallway | 0.507 | 0.395 | 0.503 | 0.394 | ||||||
| Patient to nurse station | 0.282 | 0.158 | 0.248 | 0.123 | ||||||
| Nurse station to patient | 0.262 | 0.139 | 0.229 | 0.105 | ||||||
| (Two-bed-occupied) | ||||||||||
| Room “D” | ||||||||||
| Patient in room | 0.722 | 0.655 | 0.677 | 0.597 | 0.620 | 0.536 | ||||
| Entry | 0.634 | 0.537 | 0.590 | 0.488 | 0.376 | 0.257 | ||||
| Door | 0.506 | 0.391 | 0.438 | 0.317 | 0.192 | 0.095 | ||||
| Patient to nurse station | 0.206 | 0.090 | 0.131 | 0.041 | 0.100 | 0.041 | ||||
| Nurse station to patient | 0.321 | 0.190 | 0.246 | 0.123 | 0.163 | 0.079 | ||||
| Four-bed-occupied | ||||||||||
| Room “E” | ||||||||||
| Patient in room | 0.604 | 0.511 | ||||||||
| Patient in room | 0.484 | 0.371 | ||||||||
| Patient in room | 0.444 | 0.327 | ||||||||
| Door | 0.291 | 0.187 | ||||||||
| Patient to nurse station | 0.011 | 0.001 | ||||||||
| Nurse station to patient | 0.086 | 0.023 | ||||||||
Figure 1Standard unit_Room A_standard curtain with alarm
Figure 2Standard unit_Room A_Hush Curtain with alarm
Speech Intelligibility Index (SII) and Articulation Index (AI) measures made in three patient rooms on the standard unit in three acoustic conditions: No curtain, standard curtain, and Hush Curtain®
| Refurbished unit two | No curtain | Standard curtain | Hush Curtain® | |||
|---|---|---|---|---|---|---|
| SII | AI | SII | AI | SII | AI | |
| Doorway | 0.581 | 0.487 | 0.553 | 0.454 | ||
| 2nd patient bed | 0.536 | 0.428 | 0.503 | 0.389 | ||
| Hallway | 0.357 | 0.241 | 0.442 | 0.333 | ||
| Patient to nurse station | 0.009 | 0.000 | 0.010 | 0.000 | ||
| Nurse station to patient | 0.146 | 0.039 | 0.131 | 0.034 | ||
| Two-bed-unoccupied | ||||||
| Room “G” | ||||||
| Doorway | 0.706 | 0.639 | 0.674 | 0.597 | 0.511 | 0.404 |
| 2nd patient bed | 0.592 | 0.492 | 0.570 | 0.470 | 0.375 | 0.264 |
| Hallway | 0.501 | 0.408 | 0.468 | 0.366 | 0.378 | 0.263 |
| Patient to nurse station | 0.020 | 0.003 | 0.019 | 0.001 | 0.010 | 0.000 |
| Nurse station to patient | 0.193 | 0.075 | 0.172 | 0.061 | 0.115 | 0.033 |
| One-bed-unoccupied | ||||||
| Room “H” | ||||||
| Doorway | 0.644 | 0.559 | 0.614 | 0.523 | 0.612 | 0.521 |
| Hallway | 0.449 | 0.328 | 0.403 | 0.278 | 0.394 | 0.266 |
| Patient to nurse station | 0.262 | 0.149 | 0.249 | 0.136 | 0.203 | 0.092 |
| Nurse station to patient | 0.274 | 0.150 | 0.261 | 0.135 | 0.215 | 0.096 |
Figure 3Refurbished unit_Room G_standard curtain with alarm
Figure 4Refurbished unit_Room G_Hush Curtain with alarm