| Literature DB >> 26778646 |
Sara Bresciani1, Anna Miranti2, Amalia Di Dia2, Angelo Maggio2, Christian Bracco2, Matteo Poli2, Davide Di Spirito2, Pietro Gabriele3, Michele Stasi2.
Abstract
The gamma index pass rate (%GP) of 384 helical Tomotherapy pre-patient quality assurance, acquired with ArcCHECK, is presented, analyzed, and correlated to plan characteristics. Average %GP was higher than 90% and correlated strongly with gamma method, irradiated length, pitch, maximum dose to diodes, and dose per fraction.Entities:
Keywords: Gamma index metric; Quality assurance; Tomotherapy
Mesh:
Year: 2016 PMID: 26778646 DOI: 10.1016/j.radonc.2015.12.021
Source DB: PubMed Journal: Radiother Oncol ISSN: 0167-8140 Impact factor: 6.280