Literature DB >> 26777551

Five-Year Experience of Vitamin E-Diffused Highly Cross-Linked Polyethylene Wear in Total Hip Arthroplasty Assessed by Radiostereometric Analysis.

Audrey K Nebergall1, Anders Troelsen2, Harry E Rubash1, Henrik Malchau1, Ola Rolfson3, Meridith E Greene1.   

Abstract

BACKGROUND: Vitamin E-diffused highly cross-linked polyethylene (VEPE) was developed to reduce oxidation without compromising mechanical strength. The purpose of this study was to evaluate VEPE in vivo using radiostereometric analysis (RSA) and patient-reported outcome measures (PROMs).
METHODS: Fifty-one hips were enrolled. Each patient received a VEPE liner, a porous titanium shell, and an uncemented stem with a 32-mm cobalt-chrome femoral head. Tantalum beads were inserted into the VEPE to measure femoral head penetration using RSA. RSA radiographs and PROMs were obtained preoperatively immediately after surgery, 6 months, 1, 2, 3, and 5 years after surgery.
RESULTS: Forty-seven hips returned at 3 years, and 42 hip at 5 years. The mean ± standard error of the mean proximal head penetration into the polyethylene was 0.06 ± 0.01 at 5 years. The amount of head penetration did not change significantly with increasing time in vivo. The mean ± standard error of the mean Harris Hip Score was 58 ± 2 preoperatively, which improved significantly to 93 ± 2 at 5 years (P < .001).
CONCLUSION: The head penetration into VEPE liners was low compared with non-VEPE at 5 years. After settling of the liners in the early period, no significant head penetration occurred from 2- to 5-year follow-up. All PROMs improved significantly from preoperative to postoperative and remained very favorable at 5 years. This study documents the longest-term evaluation of in vivo wear performance of VEPE.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  patient-reported outcome measures; polyethylene; radiostereometric analysis; total hip arthroplasty; vitamin E

Mesh:

Substances:

Year:  2015        PMID: 26777551     DOI: 10.1016/j.arth.2015.12.023

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

Review 1.  A prospective osseointegration retrieval analysis of second generation cementless shells.

Authors:  M Curtin; E Pomeroy; M Grigoras; T Murphy; Fiachra E Rowan
Journal:  J Clin Orthop Trauma       Date:  2022-06-03

2.  Comparison of Polyethylene Wear before and after Hip Revision with Liner Exchange Fixed with the Original Locking Mechanism.

Authors:  Xinfeng Gu; Jie He; Yiwen Tang; Yuxin Zheng
Journal:  PLoS One       Date:  2016-12-09       Impact factor: 3.240

Review 3.  Ultra-High Molecular Weight Polyethylene: Influence of the Chemical, Physical and Mechanical Properties on the Wear Behavior. A Review.

Authors:  Pierangiola Bracco; Anuj Bellare; Alessandro Bistolfi; Saverio Affatato
Journal:  Materials (Basel)       Date:  2017-07-13       Impact factor: 3.623

  3 in total