Literature DB >> 26764338

Safety of Lumbar Spine Radiofrequency Procedures in Patients Who Have Posterior Spinal Hardware.

Tim J Lamer1, Jolene Smith2, Bryan C Hoelzer3, William D Mauck3, Wenchun Qu4, Halena M Gazelka3.   

Abstract

OBJECTIVE: The purpose of this study was to determine whether the energy generated by an active radiofrequency (RF) cannula adjacent to lumbar spine hardware could result in heating of the hardware.
DESIGN: Prospective study.
SETTING: Tertiary care medical center.
SUBJECTS: Six patients with lumbar facet joint pain at the level adjacent to lumbar spine fusion hardware were studied.
METHODS: A total of 10 lumbar medial branch nerve radiofrequency lesion procedures were performed on six patients. A temperature probe was placed on the fusion hardware to continuously monitor the temperature of the hardware throughout the RF procedure.
RESULTS: The temperature of the fusion hardware increased in six of the 10 RF lesion procedures. During two of the procedures, the temperature rose rapidly to 42°C, at which time the procedure was ceased at that level.
CONCLUSION: This study demonstrated that radiofrequency lesioning to treat symptomatic facet joint pain in patients who have adjacent posterior lumbar fusion hardware may result in heat energy being transferred to the adjacent hardware. This may increase the risk of injury to the patient. Monitoring for a temperature increase is easily accomplished by inserting a temperature probe onto the surface of the hardware.
© 2016 American Academy of Pain Medicine. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.

Entities:  

Keywords:  Facet Joint Pain; Radiofrequency Neurotomy; Radiofrequency Neurotomy Safety

Mesh:

Year:  2016        PMID: 26764338     DOI: 10.1093/pm/pnv078

Source DB:  PubMed          Journal:  Pain Med        ISSN: 1526-2375            Impact factor:   3.750


  6 in total

1.  Consensus practice guidelines on interventions for cervical spine (facet) joint pain from a multispecialty international working group.

Authors:  Robert W Hurley; Meredith C B Adams; Meredith Barad; Arun Bhaskar; Anuj Bhatia; Andrea Chadwick; Timothy R Deer; Jennifer Hah; W Michael Hooten; Narayan R Kissoon; David Wonhee Lee; Zachary Mccormick; Jee Youn Moon; Samer Narouze; David A Provenzano; Byron J Schneider; Maarten van Eerd; Jan Van Zundert; Mark S Wallace; Sara M Wilson; Zirong Zhao; Steven P Cohen
Journal:  Pain Med       Date:  2021-11-26       Impact factor: 3.750

2.  Consensus practice guidelines on interventions for cervical spine (facet) joint pain from a multispecialty international working group.

Authors:  Robert W Hurley; Meredith C B Adams; Meredith Barad; Arun Bhaskar; Anuj Bhatia; Andrea Chadwick; Timothy R Deer; Jennifer Hah; W Michael Hooten; Narayan R Kissoon; David Wonhee Lee; Zachary Mccormick; Jee Youn Moon; Samer Narouze; David A Provenzano; Byron J Schneider; Maarten van Eerd; Jan Van Zundert; Mark S Wallace; Sara M Wilson; Zirong Zhao; Steven P Cohen
Journal:  Reg Anesth Pain Med       Date:  2021-11-11       Impact factor: 6.288

3.  Consensus practice guidelines on interventions for lumbar facet joint pain from a multispecialty, international working group.

Authors:  Steven P Cohen; Arun Bhaskar; Anuj Bhatia; Asokumar Buvanendran; Tim Deer; Shuchita Garg; W Michael Hooten; Robert W Hurley; David J Kennedy; Brian C McLean; Jee Youn Moon; Samer Narouze; Sanjog Pangarkar; David Anthony Provenzano; Richard Rauck; B Todd Sitzman; Matthew Smuck; Jan van Zundert; Kevin Vorenkamp; Mark S Wallace; Zirong Zhao
Journal:  Reg Anesth Pain Med       Date:  2020-04-03       Impact factor: 6.288

4.  The Efficacy of Radiofrequency Ablation for Pain Management in Patients with Pre-Existing Hardware at the Site of Ablation.

Authors:  Alaa Abd-Elsayed; Meghan Hughes; Emily Narel; Michael D Loebertman
Journal:  Pain Ther       Date:  2020-10-08

5.  Impact of radiofrequency ablation (RFA) on bone quality in a murine model of bone metastases.

Authors:  Soroush Ghomashchi; Cari M Whyne; Tricia Chinnery; Fayez Habach; Margarete K Akens
Journal:  PLoS One       Date:  2021-09-08       Impact factor: 3.240

6.  In vivo measurements of medial branch nerve depth and adjacent osseous structures for ablation of facet-related back pain: Predictors for patient candidacy.

Authors:  Hannah Zwiebel; Ron Aginsky; Arik Hananel; Daniel Baldor; Michael Gofeld; Jean-Francois Aubry; Suzanne D LeBlang
Journal:  N Am Spine Soc J       Date:  2020-08-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.