Literature DB >> 26759071

Accurate quantification of creatinine in serum by coupling a measurement standard to extractive electrospray ionization mass spectrometry.

Keke Huang1, Ming Li2, Hongmei Li2, Mengwan Li2, You Jiang2, Xiang Fang2.   

Abstract

Ambient ionization (AI) techniques have been widely used in chemistry, medicine, material science, environmental science, forensic science. AI takes advantage of direct desorption/ionization of chemicals in raw samples under ambient environmental conditions with minimal or no sample preparation. However, its quantitative accuracy is restricted by matrix effects during the ionization process. To improve the quantitative accuracy of AI, a matrix reference material, which is a particular form of measurement standard, was coupled to an AI technique in this study. Consequently the analyte concentration in a complex matrix can be easily quantified with high accuracy. As a demonstration, this novel method was applied for the accurate quantification of creatinine in serum by using extractive electrospray ionization (EESI) mass spectrometry. Over the concentration range investigated (0.166 ~ 1.617 μg/mL), a calibration curve was obtained with a satisfactory linearity (R(2) = 0.994), and acceptable relative standard deviations (RSD) of 4.6 ~ 8.0% (n = 6). Finally, the creatinine concentration value of a serum sample was determined to be 36.18 ± 1.08 μg/mL, which is in excellent agreement with the certified value of 35.16 ± 0.39 μg/mL.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26759071      PMCID: PMC4725369          DOI: 10.1038/srep19283

Source DB:  PubMed          Journal:  Sci Rep        ISSN: 2045-2322            Impact factor:   4.379


Renal failure (known as kidney failure) has received significant attention since even moderate reduction in kidney function is associated with high rate of mortality1. Early diagnosis of kidney disease, followed by appropriate medical treatments, can prevent or postpone kidney failure. Creatinine, a byproduct of muscle metabolism, is removed from the blood chiefly by the kidneys. The serum creatinine level is the most widely used and commonly accepted measure of renal function in clinical medicine23. Many techniques including electrochemical sensor4, Raman spectroscopy5, spectrophotometry67, capillary electrophoresis89, high performance liquid chromatography10, are available for the determination of creatinine in serum. Alternatively, hyphenated mass spectrometric methods including gas chromatography-mass spectrometry (GC/MS)1112, liquid chromatography-mass spectrometry (LC/MS)12131415 have been developed for the accurate quantification of creatinine in serum. Among these techniques, liquid chromatography-isotope dilution mass spectrometry (LC-IDMS) was recommended as the reference technique by National Kidney Disease Education Program16. However, traditional MS method suffers from the need of tedious, time- and cost-consuming sample pretreatment steps (derivatization, extraction, chromatographic separation, etc.) before complex raw samples are analysed. Recently introduced ambient ionization techniques, including desorption electrospray ionization (DESI)171819, direct analysis in real time (DART)20, desorption atmospheric pressure chemical ionization (DAPCI)21, dielectric barrier discharge ionization (DBDI)22, low temperature plasma probe (LTP)23, easy ambient sonic-spray ionization (EASI)24, and extractive electrospray ionization (EESI)252627282930, allow direct desorption/ionization of chemicals in raw samples under ambient conditions. Ambient MS methods require no or minimal sample preparation and tolerate chemical contamination of an ion source. It is notable that DESI was adopted for the determination of urine creatinine level19. Nonetheless, current obstacle on the quantitative analysis of ambient mass spectrometry is the matrix effect31, especially in the case that samples have high complex matrix and high quantitative accuracy is required32. For example, the signal of protonated analytes will dramatically decrease in the case: i) the proton affinities (PA) of the species in matrix are larger than PA of the analyte, ii) alkali-metals (e.g. Na+, K+) in the matrix cause the formation of alkali-metal adducts of analyte. Therefore, a standard addition method or/and isotope dilution MS were coupled to ambient ionization technique to improve the analytical accuracy3334. In this case, samples were prepared by spiking a series of stock solutions before being analyzed. For each sample analysis, at least five spiked samples had to be prepared, and consequently at least five MS analysis had to be performed3334. It is still tedious, time- and cost-consuming to prepare and to analyse the spiked samples, especially for the high throughput analysis. In this study, a simple, rapid ambient MS method was developed for accurate analysis of creatinine in serum by coupling a measurement standard to EESI.

Results

EESI-MS spectra of creatinine

To optimize the parameters of the instrument, a creatinine standard solution (0.1 μg/mL) was introduced into EESI source and a typical mass spectrum of creatinine was recorded (see Figure S1). The signal at m/z 114 corresponds to protonated creatinine molecules. The product ion spectrum of the mass selected ions of m/z 114 (inset of Figure S1) generated ionic fragments of m/z 86 by the neutral loss of CO35. A serum sample (GBW 09170) diluted 50 times with water was directly analysed by EESI-MS, and the corresponding mass spectrum was recorded (see Fig. 1). The signal at m/z 114 corresponds to protonated creatinine molecules, which was confirmed by tandem mass spectrum (inset of Fig. 1). The signals at m/z 61, 83, 136 correspond to protonated urea, sodiated urea, and sodiated creatinine, respectively. Actually, sodium ion adducts are very common in MS36, and sodiated creatinine was also observed in ambient ionization mass spectrometry such as DESI19, EESI30, and contactless atmospheric pressure ionization37.
Figure 1

A typical EESI mass spectrum of a serum sample with 50 times dilution (inset shows an MS2 spectrum of m/z 114).

Demonstration of the matrix effect

A series of creatinine standard solutions prepared with pure water were used to establish a calibration curve (see Figure S2). Afterwards, 50 fold dilution of GBW09171 was analysed by using EESI-MS. The creatinine level in the original serum sample was determined to be 13.96 ± 0.42 μg/mL, which is significantly lower than the certified value of 35.16 ± 0.39 μg/mL15. This is because matrix effect still exists during the ionization process even the serum sample was diluted 50 times with water. In addition, distribution of ion current between protonated creatinine (m/z 114) and sodiated creatinine (m/z 136) can also be a source of error.

Establishment of a calibration curve with a matrix reference material

For reliable, accurate determination of creatinine level in serum sample, a matrix reference material, which is a particular form of measurement standard, was coupled to EESI technique. GBW09170 was chosen for constructing a calibration curve as it has a lower creatinine level. Five standard solutions, prepared with the matrix reference material, were analysed by using EESI-MS in MS2 mode. The intensity of quantitative ions of m/z 86, derived from the parent ions of m/z 114, was plotted as a function of creatinine concentration (see Fig. 2). Over the concentration range investigated (0.166 ~ 1.617 μg/mL), a calibration curve was obtained with a satisfactory linearity (R2 = 0.994), and acceptable relative standard deviations (RSD) of 4.6 ~ 8.0% (n = 6). The regression equation of the calibration curve was given as y = 95.5x + 2.93. The limit of quantification (S/N = 10) was calculated to be 0.03 μg/mL (i.e. 1.5 μg/mL in original serum sample). So the method reported here can be adopted for analysing the serum samples with creatinine level ranging 1.5 ~ 80.85 μg/mL, which is a suitable range for clinical application38. To assess the inter-day variation of the response, five calibration curves were established in five consecutive days with the same series of standard solutions (see Figure S3). The relative standard deviation of five slopes was calculated to be 0.49%, however it is suggested that a calibration curve should be re-established each time the EESI source is re-assembled. Here, the standard solutions were prepared with a matrix reference material, which has similar chemical characteristics to the samples being tested15. Therefore, this calibration curve can be used to accurately interpolate the creatinine level in serum samples.
Figure 2

Calibration curve for determination of creatinine in serum with a transition ion pair m/z 114-86 as quantitative ions (Error bars designate the standard deviation, n = 6).

Quantification of creatinine in serum samples

A serum sample (GBW 09171) diluted 50 times with water was directly analysed by using EESI. The signal intensity in MS2 mode was recorded and the creatinine level in the original sample was calculated to be 36.18 ± 1.08 μg/mL. This value is in excellent agreement with the certified value of 35.16 ± 0.39 μg/mL, which was confirmed by an international comparison15. To assess the inter-assays variation, the same sample was analysed after one week, and the result was calculated to be 35.83 ± 1.01 μg/mL. The relative deviation between the results obtained from two different days is less than 0.5%. To further confirm the accuracy of this novel method, four serum samples were quantified via the proposed method and a reference method (i.e. LC-IDMS) developed previously in our lab15. The relative deviations of the results from these two methods were calculated to be 0.99–3.26%, as listed in Table 1.
Table 1

Comparison of creatinine level in serum by using LC-IDMS and EESI-MS.

Sample numberconc. by LC-IDMS (μg/mL)conc. by EESI-MS (μg/mL)Relative deviation (%)
17.677.423.26
29.038.852.00
322.5223.192.98
433.3432.010.99

Discussion

Matrix effects are a current obstacle on the quantitative analysis of ambient mass spectrometry because of ion suppression, and distribution of ion current between protonated and sodiated molecules. To resolve this problem, a matrix reference material, which is considered as a “micro-ruler” to check the accuracy of tests and analytical procedures, was coupled to EESI technique. As a demonstration, this method was applied for the accurate quantification of creatinine in serum. The results show that the accurate analyte concentration in a complex matrix can be easily obtained with only one analysis for each sample by coupling a measurement standard to EESI.

Methods

Instrumentation and Working Conditions

All the experiments were performed using a Thermo Finnigan LTQ mass spectrometer (San Jose, CA, USA) equipped with a homemade EESI source, which is described elsewhere242526272829. A schematic of EESI source is shown in Fig. 3. A methanol/water/formic acid (v/v/v, 50:50:0.1) solution was injected into channel 1 using a syringe pump with a rate of 5 μL/min and nebulized with a sheath gas (N2). A high voltage of 4 kV was applied to channel 1 to generate the primary ions. Samples or standard solutions were injected into channel 2 using another syringe pump with a rate of 5 μL/min and also nebulized with a sheath gas (N2). The neutral analytes were ionized when intersecting with the primary ions. The angle (α) formed between the channel 1 and the channel 2 was 60°. The angle (β) formed between the channel 1 and the MS inlet was 150°. The distance (a) between the emitters of these two channels was 2 mm. The distance (b) between the EESI source and the MS inlet was 2.5 cm. The temperature of the ion entrance capillary was maintained at 390 °C during the experiment. For tandem mass spectrometry, the precursor ions of interest were isolated with a window width of 1.5 mass/charge units (full-width) and helium was used as a collisional gas. MS2 experiments were performed with an activation time of 40 ms and collision energy of 24% (arbitrary unit). All the mass spectra were recorded using Xcalibur® software. The mean of the mass spectra accumulated for 1.5 min was recorded as one measurement result. An average of 6 measurements for one sample was used as the final result.
Figure 3

Schematic of the EESI source coupling with a measurement standard.

Statement

All experimental protocols were approved by the ethics committee of National Institute of Metrology, China and adhered to the tenets of the Declaration of Helsinki. Additionally, the benefits and risks of this study were clearly presented to the volunteers, and thereafter the written consent was obtained if they agreed to join the study.

Samples Preparation

Methanol (HPLC grade), and formic acid were purchased from Chinese Chemical Reagent Co., Ltd. (Shanghai, China). Ultrapure water (18.2 MΩ·cm−1) was prepared with Milli-Q Direct water purification system (Millipore, USA). Pure creatinine (SRM914a) with a purity of 99.7% was purchased from national institute of standards and technology (NIST), USA. Serum matrix reference materials GBW 09170 and GBW 09171 with certified creatinine values of 8.30 ± 0.10 μg/mL and 35.16 ± 0.39 μg/mL, were developed in our lab (i.e. national institute of metrology, China). Four serum samples, collected from volunteers including two healthy persons and two patients with kidney disease, were provided by the academy of military medical sciences (Beijing, China). All the chemicals were used directly without any further treatment.

Additional Information

How to cite this article: Huang, K. et al. Accurate quantification of creatinine in serum by coupling a measurement standard to extractive electrospray ionization mass spectrometry. Sci. Rep. 6, 19283; doi: 10.1038/srep19283 (2016).
  32 in total

1.  Development of a liquid chromatography-mass spectrometry method for the high-accuracy determination of creatinine in serum.

Authors:  Peter Stokes; Gavin O'Connor
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2003-08-25       Impact factor: 3.205

2.  Development of a dielectric barrier discharge ion source for ambient mass spectrometry.

Authors:  Na Na; Mengxia Zhao; Sichun Zhang; Chengdui Yang; Xinrong Zhang
Journal:  J Am Soc Mass Spectrom       Date:  2007-08-02       Impact factor: 3.109

3.  Determination of serum creatinine by Jaffe method and how to calibrate to eliminate matrix interference problems.

Authors:  Vratislav Chromý; Katerina Rozkosná; Pavel Sedlák
Journal:  Clin Chem Lab Med       Date:  2008       Impact factor: 3.694

4.  Document authentication at molecular levels using desorption atmospheric pressure chemical ionization mass spectrometry imaging.

Authors:  Ming Li; Bin Jia; Liying Ding; Feng Hong; Yongzhong Ouyang; Rui Chen; Shumin Zhou; Huanwen Chen; Xiang Fang
Journal:  J Mass Spectrom       Date:  2013-09       Impact factor: 1.982

5.  Neutral desorption sampling of biological surfaces for rapid chemical characterization by extractive electrospray ionization mass spectrometry.

Authors:  Huanwen Chen; Renato Zenobi
Journal:  Nat Protoc       Date:  2008       Impact factor: 13.491

6.  Direct quantification of creatinine in human urine by using isotope dilution extractive electrospray ionization tandem mass spectrometry.

Authors:  Xue Li; Xiaowei Fang; Zhiqiang Yu; Guoying Sheng; Minghong Wu; Jiamo Fu; Huanwen Chen
Journal:  Anal Chim Acta       Date:  2012-08-30       Impact factor: 6.558

7.  Easy ambient sonic-spray ionization mass spectrometry combined with thin-layer chromatography.

Authors:  Renato Haddad; Humberto M S Milagre; Rodrigo Ramos Catharino; Marcos N Eberlin
Journal:  Anal Chem       Date:  2008-03-11       Impact factor: 6.986

Review 8.  Creatinine: a review.

Authors:  S Narayanan; H D Appleton
Journal:  Clin Chem       Date:  1980-07       Impact factor: 8.327

9.  Quantitative detection of nitric oxide in exhaled human breath by extractive electrospray ionization mass spectrometry.

Authors:  Susu Pan; Yong Tian; Ming Li; Jiuyan Zhao; Lanlan Zhu; Wei Zhang; Haiwei Gu; Haidong Wang; Jianbo Shi; Xiang Fang; Penghui Li; Huanwen Chen
Journal:  Sci Rep       Date:  2015-03-04       Impact factor: 4.379

10.  Automatic sampling and analysis of organics and biomolecules by capillary action-supported contactless atmospheric pressure ionization mass spectrometry.

Authors:  Cheng-Huan Hsieh; Anil Kumar Meher; Yu-Chie Chen
Journal:  PLoS One       Date:  2013-06-06       Impact factor: 3.240

View more
  2 in total

1.  Nitrogen-terminated silicon nanoparticles obtained via chemical etching and passivation are specific fluorescent probes for creatinine.

Authors:  Lei Meng; Chengwu Lan; Zhonghu Liu; Jian-Hang Yin; Na Xu
Journal:  Mikrochim Acta       Date:  2019-05-29       Impact factor: 5.833

2.  Preparation of Isotopically Labelled Standards of Creatinine Via H/D Exchange and Their Application in Quantitative Analysis by LC-MS.

Authors:  Remigiusz Bąchor; Andrzej Konieczny; Zbigniew Szewczuk
Journal:  Molecules       Date:  2020-03-26       Impact factor: 4.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.