Dosimetric characteristics of brass mesh as bolus under megavoltage photon irradiation.

Neil D Richmond1, Jim M Daniel1, Joe R Whitbourn1, Andrew D Greenhalgh1.   

Abstract

OBJECTIVE: This article presents a set of dosimetric measurements describing the properties of brass mesh (Whiting and Davis, Attleboro Falls, MA) under megavoltage photon irradiation conditions, with particular relevance to its use in breast radiotherapy.
METHODS: The effectiveness of brass mesh as a bolus material was investigated using 6-, 15- and 6-MV flattening filter-free photon beams. The effect on dose build-up at the entrance surface, build-down at the beam-exit surface, dose with surface entrance obliquity, beam profiles, penumbra and percentage depth doses were investigated.
RESULTS: One layer of the brass mesh produces a build-up effect equivalent to 1.1 mm of water at 6 MV and 1.9 mm at 15 MV. The brass generates a backscattered component of dose, if the photon beam exits through it. Percentage depth-dose curves are largely unaffected by the mesh and are shown to be equivalent to plain-field data. Beam penumbra and profiles are unchanged by the brass except within the first millimetre (mm) of phantom, where a periodic pattern of dose enhancement is seen.
CONCLUSION: The data presented demonstrate that one layer of brass mesh provides a similar dose build-up effect equivalent to only a few millimetres of water. However, backscatter from the high atomic number (Z) mesh, at the beam exit, contributes appreciably to the overall dose surface enhancement. This dosimetric consequence cannot be neglected and indeed should be considered and accounted for, when determining the bolus effect of the brass mesh in the case of tangential breast irradiation. Advance in knowledge: This article provides dosimetric data necessary for the introduction of brass mesh bolus into the clinical setting for external-beam breast radiotherapy.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26756812     DOI: 10.1259/bjr.20150796

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  3 in total

1.  The accuracy of treatment planning system dose modelling in the presence of brass mesh bolus.

Authors:  Neil Richmond
Journal:  Rep Pract Oncol Radiother       Date:  2017-07-22

2.  Dosimetric characteristics of a thin bolus made of variable shape tungsten rubber for photon radiotherapy.

Authors:  Katsuya Okuhata; Mikoto Tamura; Hajime Monzen; Yasumasa Nishimura
Journal:  Phys Eng Sci Med       Date:  2021-09-20

3.  Surface Dose Measurements in Chest Wall Postmastectomy Radiotherapy to Achieve Optimal Dose Delivery with 6 MV Photon Beam.

Authors:  Dilson Lobo; Sourjya Banerjee; Challapalli Srinivas; M S Athiyamaan; Shreyas Reddy; Johan Sunny; Ramamoorthy Ravichandran; Himani Kotian; P U Prakash Saxena
Journal:  J Med Phys       Date:  2021-11-20
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.